U.S. Nuclear Regulatory Commission Operations Center Event Reports For 01/12/2016 - 01/13/2016 ** EVENT NUMBERS ** | Agreement State | Event Number: 51636 | Rep Org: NEW JERSEY DEPT OF ENVIR PROTECTION Licensee: ENGLEWOOD HOSPITAL & MEDICAL CENTER Region: 1 City: ENGLEWOOD State: NJ County: License #: 332543 Agreement: Y Docket: NRC Notified By: RICHARD PEROS HQ OPS Officer: DANIEL MILLS | Notification Date: 01/05/2016 Notification Time: 16:13 [ET] Event Date: 01/05/2016 Event Time: [EST] Last Update Date: 01/05/2016 | Emergency Class: NON EMERGENCY 10 CFR Section: AGREEMENT STATE | Person (Organization): ANTHONY DIMITRIADIS (R1DO) NMSS_EVENTS_NOTIFIC (EMAI) | Event Text AGREEMENT STATE REPORT - MEDICAL UNDERDOSE The following was received from New Jersey via fax: "The hospital's Radiation Safety Officer reported that a patient was due to receive a therapeutic treatment of Y-90 SIR-Spheres. The prescribed dose was 10.5 milliCuries to the left lobe of the liver. The dose administered was only 65% of the prescribed dose. Approximately 3.2 milliCuries remained in the vial after the administration was completed. The patient was underdosed and has been informed. Additional details will be forthcoming from the hospital." A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient. | Power Reactor | Event Number: 51646 | Facility: CALLAWAY Region: 4 State: MO Unit: [1] [ ] [ ] RX Type: [1] W-4-LP NRC Notified By: TODD WITT HQ OPS Officer: VINCE KLCO | Notification Date: 01/12/2016 Notification Time: 11:14 [ET] Event Date: 01/11/2016 Event Time: 12:00 [CST] Last Update Date: 01/12/2016 | Emergency Class: NON EMERGENCY 10 CFR Section: 26.719 - FITNESS FOR DUTY | Person (Organization): RAY KELLAR (R4DO) | Unit | SCRAM Code | RX CRIT | Initial PWR | Initial RX Mode | Current PWR | Current RX Mode | 1 | N | Y | 100 | Power Operation | 100 | Power Operation | Event Text FITNESS FOR DUTY "Event Report per 10 CFR 26.719(b)(2)(ii) "On January 11, 2016, Callaway determined a violation of two provisions of the site Fitness For Duty policy were committed offsite by a non-licensed supervisory employee. Unescorted access for the employee has been denied." The licensee notified the NRC Resident Inspector. | Power Reactor | Event Number: 51649 | Facility: CALLAWAY Region: 4 State: MO Unit: [1] [ ] [ ] RX Type: [1] W-4-LP NRC Notified By: DAVID LANTZ HQ OPS Officer: JEFF ROTTON | Notification Date: 01/12/2016 Notification Time: 18:37 [ET] Event Date: 01/12/2016 Event Time: 12:53 [CST] Last Update Date: 01/12/2016 | Emergency Class: NON EMERGENCY 10 CFR Section: 50.72(b)(3)(xiii) - LOSS COMM/ASMT/RESPONSE | Person (Organization): RAY KELLAR (R4DO) | Unit | SCRAM Code | RX CRIT | Initial PWR | Initial RX Mode | Current PWR | Current RX Mode | 1 | N | Y | 100 | Power Operation | 100 | Power Operation | Event Text LOSS OF VENTILATION SYSTEM TO EMERGENCY OFF-SITE FACILITY "On January 12, 2016, during performance of an EOF [Emergency Off-Site Facility] Diesel Operability Test, the EOF Air Conditioning unit and EOF Air Return fan did not run as expected. Per plant procedure, the operators placed the system in filtration mode and then back in normal mode. Again, both units did not run as expected. The EOF was declared non-functional due to the failure of the air conditioning unit and fan. "The EOF is available for emergency response purposes unless the temperature can not be maintained or a release is in progress. The backup EOF is available for use if needed." The licensee notified the NRC Resident Inspector. | Power Reactor | Event Number: 51650 | Facility: QUAD CITIES Region: 3 State: IL Unit: [1] [2] [ ] RX Type: [1] GE-3,[2] GE-3 NRC Notified By: GARY HEUSINKVELD HQ OPS Officer: DONALD NORWOOD | Notification Date: 01/12/2016 Notification Time: 19:08 [ET] Event Date: 01/12/2016 Event Time: 13:40 [CST] Last Update Date: 01/12/2016 | Emergency Class: NON EMERGENCY 10 CFR Section: 50.72(b)(3)(v)(C) - POT UNCNTRL RAD REL | Person (Organization): ERIC DUNCAN (R3DO) | Unit | SCRAM Code | RX CRIT | Initial PWR | Initial RX Mode | Current PWR | Current RX Mode | 1 | N | Y | 100 | Power Operation | 100 | Power Operation | 2 | N | Y | 100 | Power Operation | 100 | Power Operation | Event Text MOMENTARY LOSS OF SECONDARY CONTAINMENT "On January 12, 2016, at 1340 CST, an alarm was received indicating secondary containment differential pressure rose unexpectedly above the Technical Specification Surveillance Requirement, SR 3.6.4.1.1, limit of 0.10 inch of vacuum water gauge. A reactor building supply fan was immediately secured to restore differential pressure below the Technical Specification limit. The secondary containment differential pressure returned to below the Technical Specification limit within one minute. The Standby Gas Treatment System remained in standby, and fully operable. At 1341 CST, the secured reactor building supply fan was restarted to restore the normal ventilation lineup. The secondary containment differential pressure remained below the Technical Specification limit. Troubleshooting is in progress. "This condition represents a failure to meet Surveillance Requirement 3.6.4.1.1. As a result, entry into Technical Specification 3.6.4.1, Condition A, was made momentarily due to secondary containment being inoperable. "Given the temporary loss of secondary containment, this event is reportable under 50.72(b)(3)(v)(c) as an event or condition that could have prevented the fulfillment of a safety function. "The NRC Senior Resident Inspector has been notified." | |