Part 21 Report - 1998-230
ACCESSION #: 9805080133
| ELECTRONIC TECHNOLOGIES Analytical Instruments An Allegheny Teledyne Company 16830 CHESTNUT STREET, CITY OF INDUSTRY, CA 91748-1020 P.O. BOX 1580, CITY OF INDUSTRY, CA 91749-1580 TELEPHONE: (626) 934-1500 OR (626) 961-9221 FAX: (626) 961-2538 |
February 18, 1998
U.S. Nuclear Regulatory Commission, Region V
1450 Maria Lane, Suite 210
Walnut Creek, CA 94596
To Whom it May Concern:
This letter serves as Teledyne Electronic Technologies/Analytical Instruments(TET/AI) corrective action plan notification to the Nuclear Regulatory Commission (NRC) related to the problem Teledyne informed you of on November 25, 1997 with switch part number S505. The attachments to this letter have been sent to all of the customers listed at the bottom of the letter addressed to "Dear Customer".
| Sincerely,
| ||
| Brian E. Greene Quality Assurance Director | Robert K. Johns Nuclear Project Engineer | |
| Reference #:BG02-98-016 | ||
| ELECTRONIC TECHNOLOGIES Analytical Instruments An Allegheny Teledyne Company 16830 CHESTNUT STREET, CITY OF INDUSTRY, CA 91748-1020 P.O. BOX 1580, CITY OF INDUSTRY, CA 91749-1580 TELEPHONE: (626) 934-1500 OR (626) 961-9221 FAX: (626) 961-2538 | ||
Dear Customer:
We at Teledyne Electronic Technologies/Analytical Instruments are writing to inform you of a defect we have become aware of in two switches that are part of the configuration of the Model 225 Hydrogen Containment Monitor. The defective switches are Teledyne part numbers S505 and S523.
We received a report from Niagara Mohawk Power Corporation which describes a problem related to the S505 switch. We have been informed by the switch manufacturer that all switches, Teledyne part number S505 and S523, with date codes from 9204 through 9336 may potentially have a lubrication problem.
At this point we have notified the NRC region V office of the issue and have submitted the attached documents to explain the cause of the problem and the corrective action plan Teledyne is recommending.
If you have any questions or need further clarification, please call and ask for Brian Greene, Quality Assurance Director or Kim Johns, Nuclear Engineer at (888) 798-8168.
| CC: | |
| USNRC Region V Duquesne Light & Power Carolina Power & Light Florida Power Corp. Korea Electric Co. Duke Power Co. Vermont Yankee Nuclear Power Corp. Hidroelectrica Espanola, S.A. Niagara Mohawk Power Corp. | General Electric Co. Commonwealth Edison Co. Bernische Kraftwerks AG Westinghouse Nuclear Espanola Louisiana Power & Light Taiwan-Power Co. Perfect Technologies LTD ACAL Auriema LTD Nuclenor, Spain |
Reference #: BG02-98-012
Please make all entries legible
CORRECTIVE/PREVENTIVE ACTION REQUEST
| Date: | No. | |
| Initiator/Employee | S/O or W/O# (if applicable) | |
| Brian E. Greene | N/A | Due Dates: |
| Plan: 1/6/98 Completion: 2/3/98 Extension: | ||
| Program/Project/Customer | Part Description/Part No. | Responsible Dept. |
| Niagara Mohawk Power Corp. | Switch/S505 | Eng./QA |
| Actionee assigned | Date | QA Manager or Designee | Date |
DESCRIPTION OF PROBLEM/CONDITION
In a panel walkdown at the Niagara Mohawk plant it was noticed that the H2 and O2 indications were increasing on the channel 11 system. The O2 reading was above the alarm setpoint and the alarm was not received. the switch was found to be the cause of the alarm failure
CAUSE:
Defective component supplied from the vendor the Teledyne. Refer to the vendor letter date May 16, 1994.
CORRECTIVE/PREVENTIVE ACTION
Refer to the attached memorandum from Kim Johns dated February 11, 1998
PLAN ACCEPTANCE SIGN-OFF
| QA Signature | Title | Date |
| FOLLOW-UP VERIFICATION | ||
| COMPLETION SIGN-OFF | ||
| Actionee Signature | Title | Date |
| QA Signature | Title | Date |
Form 011 Rev. 1
| TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments An Allegheny Teledyne Company |
MEMORANDUM
| DATE: | February 11, 1998 |
| TO: | Brian Greene |
| FROM: | Kim Johns |
| cc: | Rick Pettit, Ed Chu, Mann Nguyen |
| SUBJECT: | Engineering evaluation of problems reported in Niagara Mohawk Power Corp. letter dated November 13, 1997 to Brian Greene. |
On November 13, 1997 a letter was received from Niagara Mohawk Power Corp. (NMPC) which stated that they were experiencing switch failures associated with the two (2) Model 225 CMA-X Containment Monitoring Systems at the Nine Mile Point Unit 1 Nuclear Station. The following is the evaluation of the reported switch (TET/AI Part No. S505) failure which did result in a 10CFR21 report filing dated November 25, 1997.
Analysis of the Suspected Defect
The switch is a miniature toggle type with momentary action. The switch is manufactured by Eaton Corporation. The reported failure is that the switch (TET/AI P/N S505) contacts failed in an open circuit condition causing the inlet gas flow to the system to be interrupted and the alarms to be inhibited.
NMPC engineering department reported that one switch was disassembled and the internal components were inspected and found corrosion internal to the switch, no lubricant in the switch, and the adhesive had dried up and let go.
Two (2) switches were returned by NMPC for TET/AI to evaluate. One switch was described above as being disassembled and the other was replaced but not disassembled. These two (2) switches have Eaton Corporation date codes of 9246 and were returned to the manufacture for evaluation. The manufacture also provided a copy of their Product Reliability Notification dated May 16, 1994 (see the enclosed report).
The switch manufacture confirmed that the two (2) suspect switches fall into the date code range of 9204 through 9336. The report identified a potential problem with the grease (i.e. lubricant) used in the manufacture of the switches. The report is also in agreement with the evaluation done by the NMPC engineering department.
Corrective Action Plan
TET/AI recommends that all parts (i.e. TET/AI P/N's S505 and S523) be visually inspected for the defective date codes identified in the Eaton Corporation letter dated May 16, 1994. TET/AI purchases commercial grade items and then dedicates these items as one lot. Upon receipt, TET/AI assigns a lot number. This lot number does not correlate with the vendors date codes. Because of this, TET/AI cannot provide a cross reference between the known defective manufactures date codes and the TET/AI lot numbers.
Reference #: RKJ:970211A
| TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments An Allegheny Teledyne Company | ||
| MEMORANDUM | ||
| DATE: | February 11, 1998 | |
| SUBJECT: | Switch Problems | |
| PAGE 2 OF 2 | ||
The following standard documents (attached) identify the location of the switches (i.e. TET/AI P/N's S505 and S523) of the Model 225CM Containment Monitor.
| Document Number | Description | |
| D61227 | Control Cabinet Assembly | |
| C24300 | Control Unit, H2 Module, Front Panel Sub-assembly | |
| C24301 | Control Unit, O2 Module, Front Panel Sub-assembly | |
| C28576 | Control Unit, Sequencer Module, Assembly | |
The above documents are used on standard systems. If the system is custom, the description and location of the switches would be similar.
If any switches are found with these date codes and they were purchased from TET/AI, please contact either Pat Stamper of the Customer Service Department or Brian Greene, the Quality Assurance Director, and the switches will be replaced free of charge.
Reference #: RKJ:970211A
Eaton Corporation
Commercial Controls Division
331 City Park Drive, S.E.
Arab, AL 35016
205 / 586-6061
FAX: 205 / 586-1325
EATON
URGENT! IMMEDIATE ATTENTION REQUIRED
May 16, 1994
To: Our Valued Customer
Subject: PRODUCT RELIABILITY NOTIFICATION
Analysis of product warranty claims has identified a potential problem on certain Miniature Toggle and Rocker switches produced between February 1992 and August 1993, which might create a risk of personal injury or property damage. Date codes affected are 9204 through 9336.
A potential exists for the grease, used inside these switches for mechanical wear purposes, to lose its lubricity causing intermittent continuity and/or difficulty in tripping the actuator. Based upon analysis of returned product and extensive internal evaluation, it is Eaton's opinion that the potential problem is most likely to occur under conditions of exposure to elevated temperatures or high humidity for extended periods of time. The number of warranty claims reported is small compared to the volume of product produced in this time period. To date, less than 1% of product produced has been returned to Eaton for replacement. However in some circumstances, under conditions stated above, this could possibly result in an application problem should the switch not operate or make circuit. SINCE EATON DOES NOT KNOW THE SPECIFICS OF YOUR APPLICATION, IT IS VERY IMPORTANT THAT YOU DETERMINE WHETHER THE CONDITIONS STATED ABOVE REPRESENT A RISK OF PERSONAL INJURY OR PROPERTY DAMAGE POTENTIAL IN YOUR PRODUCT.
To reduce the potential of this problem continuing; Eaton has made process changes in the preservation of mechanical components and converted to a grease containing additional anti-corrosive additives and higher operating temperature capabilities. Switches produced since September 1993 ( Date Code 9337 ) should not exhibit these potential characteristics when applied in accordance with advertised specifications.
PLEASE FOLLOW THESE VERY IMPORTANT INSTRUCTIONS:
| 1. | To aid you in identification of specific switches with the potential for characteristics as described above, we have enclosed a list of catalog numbers, ship dates, order numbers, and quantities which were shipped to you during this time period. |
| 2. | Evaluate the attached list to determine whether the potential problem described above has been experienced by your company or whether the conditions described are relevant to your application. If so, please follow return/replacement procedures which follow. |
| 3. | Suspect product from your inventory, as identified on the enclosed list should be returned to Eaton for replacement following the instructions given below. All switches returned will be replaced on a one for one basis at no charge to you. |
Return/Replacement Procedure:
| 1. | Contact the Arab Manufacturing Plant at 205-586-9220 and ask for Anthony Sorcic who will assign a Return Authorization Number. It is important to write this assigned number on the packing list (itemized by product catalog number and quantity) and on the outside of the shipping carton. | |
| 2. | Supply Anthony Sorcic with a list of material to be returned for replacement (identified by product catalog number, quantity and date code by phone 205-586-9220; or FAX 205-586-1325; or mail to: | |
| Eaton Corporation Commercial Controls Division 331 City Park Drive Arab, Alabama 35016 Return Number ____________ Attn: Anthony Sorcic | ||
| 3. | Return the suspect devices to the above address. Eaton will pay all return and reshipment costs. | |
| 4. | We request that you respond within 15 days of receipt of this notice. | |
We thank you for your help.
Ferrel Click
Quality Assurance Manager
Attachments




*** END OF DOCUMENT ***