Part 21 Report - 1996-720
GENERAL INFORMATION or OTHER EVENT NUMBER: 30845 LICENSEE: TSI NOTIFICATION DATE: 08/07/96 CITY: ST. PAU; REGION: 3 NOTIFICATION TIME: 17:45 [ET] COUNTY: STATE: MN EVENT DATE: 08/07/96 LICENSE#: AGREEMENT: N EVENT TIME: 12:00[CDT] DOCKET: LAST UPDATE DATE: 08/07/96 NOTIFICATIONS NRC NOTIFIED BY: TSI VERN HODGE NRR HQ OPS OFFICER: DOUG WEAVER EMERGENCY CLASS: NOT APPLICABLE 10 CFR SECTION: CCCC 21.21 UNSPECIFIED PARAGRAPH EVENT TEXT PARTS 21 REPORTED BY TSI, ENVIRONMENTAL MEASUREMENTS AND CONTROLS DIVISION TSI REPORTED THAT SOME OF THEIR INSTRUMENTS MAY NOT PERFORM CORRECTLY IN ELEVATED TEMPERATURE AND HIGH HUMIDITY ENVIRONMENTS. THE FOLLOWING INSTRUMENTS MAY HAVE THIS PROBLEM. VELOCICALC MODELS 8355, 8357, 8360, 8382, 8383, AND 8388 WITH SERIAL NUMBERS BEGINNING WITH THE DIGITS 604 THROUGH 606 OR 9604 THROUGH 9606 VELOCICHECK MODELS 8330, 8340 WITH SERIAL NUMBERS BEGINNING WITH THE DIGITS 9601 THROUGH 9606 ACCUBALANCE MODEL 8370 WITH SERIAL NUMBERS BEGINNING WITH THE DIGITS 601 THROUGH 606 DP-CALC MODELS 8702, 8704 WITH SERIAL NUMBERS BEGINNING WITH THE DIGITS 9604 THROUGH 9606 TSI WILL SEND THIS INFORMATION TO ALL IDENTIFIED OWNERS OF THE INSTRUMENTS LISTED ABOVE.
Page Last Reviewed/Updated Wednesday, March 24, 2021
Page Last Reviewed/Updated Wednesday, March 24, 2021