Due to a lapse in appropriations, the NRC has ceased normal operations. However, excepted and exempted activities necessary to maintain critical health and safety functions—as well as essential progress on designated critical activities, including those specified in Executive Order 14300—will continue, consistent with the OMB-Approved NRC Lapse Plan.

Event Notification Report for April 29, 2002

                         
                    U.S. Nuclear Regulatory Commission
                              Operations Center

                              Event Reports For
                           04/26/2002 - 04/29/2002

                              ** EVENT NUMBERS **

38875  38878  38879  38880  38881  

+------------------------------------------------------------------------------+
|General Information or Other                     |Event Number:   38875       |
+------------------------------------------------------------------------------+
                         
+------------------------------------------------------------------------------+
+------------------------------------------------------------------------------+
| REP ORG:  ALABAMA RADIATION CONTROL            |NOTIFICATION DATE: 04/24/2002|
|LICENSEE:  UNIVERSITY OF ALABAMA MEDICAL CENTER |NOTIFICATION TIME: 18:10[EDT]|
|    CITY:  BIRMINGHAM               REGION:  2  |EVENT DATE:        04/24/2002|
|  COUNTY:                            STATE:  AL |EVENT TIME:             [CDT]|
|LICENSE#:  266                   AGREEMENT:  Y  |LAST UPDATE DATE:  04/24/2002|
|  DOCKET:                                       |+----------------------------+
|                                                |PERSON          ORGANIZATION |
|                                                |ANNE BOLAND          R2      |
|                                                |MELVYN LEACH         NMSS    |
+------------------------------------------------+                             |
| NRC NOTIFIED BY:  BRADLEY GRINSTEAD            |                             |
|  HQ OPS OFFICER:  RICH LAURA                   |                             |
+------------------------------------------------+                             |
|EMERGENCY CLASS:          NON EMERGENCY         |                             |
|10 CFR SECTION:                                 |                             |
|NAGR                     AGREEMENT STATE        |                             |
|                                                |                             |
|                                                |                             |
|                                                |                             |
|                                                |                             |
+------------------------------------------------------------------------------+
                                   EVENT TEXT                                  
+------------------------------------------------------------------------------+
| AGREEMENT STATE REPORT OF LOST IRIDIUM-192 SOURCE                            |
|                                                                              |
| The Alabama State Radiation Control Office reported that the University of   |
| Alabama Medical Center was missing a 250-millicurie iridium-192 source.  The |
| source is used for cardiology purposes and was last seen on April 22, 2002.  |
| The hospital was conducting searches to try to locate the missing source.    |
+------------------------------------------------------------------------------+

+------------------------------------------------------------------------------+
|Power Reactor                                    |Event Number:   38878       |
+------------------------------------------------------------------------------+
                         
+------------------------------------------------------------------------------+
+------------------------------------------------------------------------------+
| FACILITY: INDIAN POINT             REGION:  1  |NOTIFICATION DATE: 04/26/2002|
|    UNIT:  [2] [] []                 STATE:  NY |NOTIFICATION TIME: 09:06[EDT]|
|   RXTYPE: [2] W-4-LP,[3] W-4-LP                |EVENT DATE:        04/26/2002|
+------------------------------------------------+EVENT TIME:        08:35[EDT]|
| NRC NOTIFIED BY:  JOHN BALLETTA                |LAST UPDATE DATE:  04/26/2002|
|  HQ OPS OFFICER:  LEIGH TROCINE                +-----------------------------+
+------------------------------------------------+PERSON          ORGANIZATION |
|EMERGENCY CLASS:          NON EMERGENCY         |CLIFFORD ANDERSON    R1      |
|10 CFR SECTION:                                 |TOM KOSHY            NRR     |
|ACOM 50.72(b)(3)(xiii)   LOSS COMM/ASMT/RESPONSE|                             |
|                                                |                             |
|                                                |                             |
|                                                |                             |
+-----+----------+-------+--------+-----------------+--------+-----------------+
|UNIT |SCRAM CODE|RX CRIT|INIT PWR|   INIT RX MODE  |CURR PWR|  CURR RX MODE   |
+-----+----------+-------+--------+-----------------+--------+-----------------+
|2     N          Y       100      Power Operation  |100      Power Operation  |
|                                                   |                          |
|                                                   |                          |
+------------------------------------------------------------------------------+
                                   EVENT TEXT                                   
+------------------------------------------------------------------------------+
| EMERGENCY RESPONSE DATA SYSTEM (ERDS) INOPERABLE DUE TO PLANNED OUTAGE OF    |
| THE PLANT PROCESS COMPUTER SYSTEM                                            |
|                                                                              |
| The following text is a portion of a facsimile received from the licensee:   |
|                                                                              |
| "On [April 26, 2002,] at 0835 hours EDT, the Indian Point Unit 2 plant ERDS  |
| was taken out of service to perform planned modifications to the plant       |
| process computer system.  ERDS will remain inoperable until approximately    |
| May 4, 2002."                                                                |
|                                                                              |
| "This report is being made in accordance with the requirements of 10 CFR     |
| 50.72(b)(3)(xiii); any event that results in a loss of offsite               |
| communications capability."                                                  |
|                                                                              |
| "The Senior Resident Inspector has been briefed on the modifications and     |
| schedule.  The NRC was previously notified of this planned outage via letter |
| dated April 4, 2002."                                                        |
+------------------------------------------------------------------------------+

+------------------------------------------------------------------------------+
|General Information or Other                     |Event Number:   38879       |
+------------------------------------------------------------------------------+
                         
+------------------------------------------------------------------------------+
+------------------------------------------------------------------------------+
| REP ORG:  OHIO BUREAU OF RADIATION PROTECTION  |NOTIFICATION DATE: 04/26/2002|
|LICENSEE:  RIVERSIDE METHODIST HOSPITAL         |NOTIFICATION TIME: 13:06[EDT]|
|    CITY:  COLUMBUS                 REGION:  3  |EVENT DATE:        04/23/2002|
|  COUNTY:                            STATE:  OH |EVENT TIME:             [EDT]|
|LICENSE#:  02120-250070          AGREEMENT:  Y  |LAST UPDATE DATE:  04/26/2002|
|  DOCKET:                                       |+----------------------------+
|                                                |PERSON          ORGANIZATION |
|                                                |JOHN JACOBSON        R3      |
|                                                |DOUGLAS BROADDUS     NMSS    |
+------------------------------------------------+                             |
| NRC NOTIFIED BY:  MARK LIGHT                   |                             |
|  HQ OPS OFFICER:  RICH LAURA                   |                             |
+------------------------------------------------+                             |
|EMERGENCY CLASS:          NON EMERGENCY         |                             |
|10 CFR SECTION:                                 |                             |
|NAGR                     AGREEMENT STATE        |                             |
|                                                |                             |
|                                                |                             |
|                                                |                             |
|                                                |                             |
+------------------------------------------------------------------------------+

                                   EVENT TEXT                                   
+------------------------------------------------------------------------------+
| AGREEMENT STATE REPORT OF A MEDICAL MISADMINISTRATION                        |
|                                                                              |
| A medical misadministration occurred on April 23, 2002 that resulted in the  |
| patient receiving a 36% overdose (i.e., 10.9 gray vice 8 gray) during        |
| brachytherapy to treat coronary vessels.  The isotope being used for the     |
| treatment is IR-192. The error was discovered on April 25, 2002 when the     |
| hospital was setting up to treat another patient.  Specifically, a wrong     |
| decay table was used in the calculations.  No adverse effect on the patient  |
| was anticipated.  The patient has been notified.                             |
|                                                                              |
| The State reported that RIII (K. NULL) has been notified.                    |
+------------------------------------------------------------------------------+

+------------------------------------------------------------------------------+
|Power Reactor                                    |Event Number:   38880       |
+------------------------------------------------------------------------------+
                         
+------------------------------------------------------------------------------+
+------------------------------------------------------------------------------+
| FACILITY: MILLSTONE                REGION:  1  |NOTIFICATION DATE: 04/26/2002|
|    UNIT:  [] [] [3]                 STATE:  CT |NOTIFICATION TIME: 16:10[EDT]|
|   RXTYPE: [1] GE-3,[2] CE,[3] W-4-LP           |EVENT DATE:        04/26/2002|
+------------------------------------------------+EVENT TIME:        15:51[EDT]|
| NRC NOTIFIED BY:  ROBERT SMITH                 |LAST UPDATE DATE:  04/26/2002|
|  HQ OPS OFFICER:  RICH LAURA                   +-----------------------------+
+------------------------------------------------+PERSON          ORGANIZATION |
|EMERGENCY CLASS:          NON EMERGENCY         |CLIFFORD ANDERSON    R1      |
|10 CFR SECTION:                                 |DAVID MATTHEWS       NRR     |
|AUNA 50.72(b)(3)(ii)(B)  UNANALYZED CONDITION   |                             |
|                                                |                             |
|                                                |                             |
|                                                |                             |
+-----+----------+-------+--------+-----------------+--------+-----------------+
|UNIT |SCRAM CODE|RX CRIT|INIT PWR|   INIT RX MODE  |CURR PWR|  CURR RX MODE   |
+-----+----------+-------+--------+-----------------+--------+-----------------+
|                                                   |                          |
|                                                   |                          |
|3     N          Y       100      Power Operation  |100      Power Operation  |
+------------------------------------------------------------------------------+
                                   EVENT TEXT                                  
+------------------------------------------------------------------------------+
| POTENTIAL FOR HOT SHORTS IN PORV CIRCUITS                                    |
|                                                                              |
| "Recent review of Station Procedures regarding fire in the Control Room and  |
| other specific fire areas reveals an inconsistency with the plant Licensing  |
| Basis. The unit Licensing Basis takes credit for prevention of spurious      |
| operation of the PORVs by removing power from the valves prior to spurious   |
| operation. Current procedures require removing power from the PORVs but does |
| not require removal of power prior to spurious operation. Spurious operation |
| of the PORVs due to a hot short is an unanalyzed event.                      |
|                                                                              |
| "Compensatory actions to minimize risk of fire in areas of concern are being |
| implemented which include increased surveillances to verify operability of   |
| detection and suppression, control of transient combustibles and ignition    |
| sources. The PORVs are considered operable.                                  |
|                                                                              |
| "This event is being reported lAW 10CFR50.72(b)(3)(ii)(B) as an event or     |
| condition that results in the Nuclear Power Plant being in an unanalyzed     |
| condition that significantly degrades plant safety."                         |
|                                                                              |
| The NRC Resident Inspector was notified.                                     |
+------------------------------------------------------------------------------+

+------------------------------------------------------------------------------+
|Power Reactor                                    |Event Number:   38881       |
+------------------------------------------------------------------------------+
                         
+------------------------------------------------------------------------------+
+------------------------------------------------------------------------------+
| FACILITY: LASALLE                  REGION:  3  |NOTIFICATION DATE:04/27/2002 |
|    UNIT:  [1] [2] []                STATE:  IL |NOTIFICATION TIME:15:51[EDT] |
|   RXTYPE: [1] GE-5,[2] GE-5                    |EVENT DATE:       04/27/2002 |
+------------------------------------------------+EVENT TIME:       11:40[CDT] |
| NRC NOTIFIED BY:  GRANLUND                     |LAST UPDATE DATE: 04/27/2002 |
|  HQ OPS OFFICER:  CHAUNCEY GOULD               +-----------------------------+
+------------------------------------------------+PERSON         ORGANIZATION  |
|EMERGENCY CLASS:          NON EMERGENCY         |JOHN JACOBSON        R3      |
|10 CFR SECTION:                                 |                             |
|APRE 50.72(b)(2)(xi)     OFFSITE NOTIFICATION   |                             |
|                                                |                             |
|                                                |                             |
|                                                |                             |
+-----+----------+-------+--------+-----------------+--------+-----------------+
|UNIT |SCRAM CODE|RX CRIT|INIT PWR|   INIT RX MODE  |CURR PWR|  CURR RX MODE   |
+-----+----------+-------+--------+-----------------+--------+-----------------+
|1     N          Y       100      Power Operation  |100      Power Operation  |
|2     N          Y       100      Power Operation  |100      Power Operation  |
|                                                   |                          |
+------------------------------------------------------------------------------+
                                   EVENT TEXT                                   
+------------------------------------------------------------------------------+
| OFFSITE NOTIFICATION WAS MADE OF A BOATING ACCIDENT ON THE LASALLE STATION   |
| COOLING LAKE                                                                 |
|                                                                              |
|                                                                              |
| On April 27, 2002, 1140 CDST the Main Control Room received a report from    |
| Outside Rounds Operator that a fishing boat had capsized on the station      |
| cooling lake, and two individuals aboard the capsized boat were at the       |
| Wastewater Treatment Facility after swimming ashore.  One individual aboard  |
| the capsized boat remains missing.  Currently the boat has been recovered,   |
| but after a search of the lake perimeter the individual remains missing.     |
| The Ottawa Search and Rescue Team was on site to perform rescue operations,  |
| but has determined that lake conditions are too rough to perform a water     |
| based rescue operation.  State Police and LaSalle County Sheriff have been   |
| notified.  Marseilles and Seneca Fire Departments are coming to the site to  |
| assist in a ground search.  Department of Natural Resources officials are    |
| also assisting with the missing person search.                               |
|                                                                              |
|                                                                              |
| The NRC Resident Inspector was notified.                                     |
+------------------------------------------------------------------------------+