Skip to main content

Event Notification Report for May 27, 2026

U.S. Nuclear Regulatory Commission
Operations Center

EVENT REPORTS FOR
05/26/2026 - 05/27/2026

EVENT NUMBERS
58287
Agreement State
Event Number: 58287
Rep Org: NJ Rad Prot And Rel Prevention Pgm
Licensee: Valley Hospital Luckow Pavillion
Region: 1
City: Paramus   State: NJ
County:
License #: 425378
Agreement: Y
Docket:
NRC Notified By: Richard Peros
HQ OPS Officer: Ian Howard
Notification Date: 05/19/2026
Notification Time: 16:51 [ET]
Event Date: 05/18/2026
Event Time: 00:00 [EDT]
Last Update Date: 05/19/2026
Emergency Class: Non Emergency
10 CFR Section:
Agreement State
Person (Organization):
Anagnostopoulos, Harold (R1DO)
NMSS_EVENTS_NOTIFICATION (EMAIL)
Event Text
AGREEMENT STATE REPORT - MEDICAL EVENT

The following information was provided by the New Jersey Department of Environmental Protection (NJDEP) via email:

"The licensee reported that during a treatment with an Elekta Leksell Gamma Knife Icon unit [on 5/18/2026], the unit halted treatment due to a collision error twice during the treatment. The licensee was able to clear the error the first time and continue with the treatment but were unable to do so the second time. This caused one of the treatment areas to only receive 7.9 Gy, compared to a planned dose of 21 Gy.

"After the licensee investigated, they believed the patient's blanket was bumping into the collimator cap, causing the collision errors. They have since tested the unit and, since it passed all testing, placed it back in service."

State Event Report ID number: TBD

A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.