Event Notification Report for June 15, 2021

U.S. Nuclear Regulatory Commission
Operations Center

EVENT REPORTS FOR
06/14/2021 - 06/15/2021

Agreement State
Event Number: 55316
Rep Org: NJ DEPT OF ENVIRONMENTAL PROTECTION
Licensee: Private Citizen
Region: 1
City: Chatham   State: NJ
County:
License #: N/A
Agreement: Y
Docket:
NRC Notified By: Richard Peros
HQ OPS Officer: Jeffrey Whited
Notification Date: 06/17/2021
Notification Time: 18:07 [ET]
Event Date: 06/16/2021
Event Time: 00:00 [EDT]
Last Update Date: 06/17/2021
Emergency Class: Non Emergency
10 CFR Section:
Agreement State
Person (Organization):
LILLIENDAHL, JON (R1)
NMSS_EVENTS_NOTIFICATION, (EMAIL)
Event Text
EN Revision Imported Date: 7/16/2021

EN Revision Text: AGREEMENT STATE REPORT - PRIVATE CITIZEN IN POSSESSION OF NUCLEAR MATERIAL

The following was received from the New Jersey Department of Environmental Protection (DEP) via email:

"On June 16, 2021, the DEP was notified that family members were cleaning out the home of a deceased relative, and discovered an item marked "radioactive." The DEP responded on June 17, 2021. The item in question appeared to be some sort of tile, or piece of rock, and was identified as containing Ra-226. There was some removable activity on this item. It was re-wrapped in the lead sheeting it was found in and placed in a plastic bucket, sealed, labelled, and secured in a safe location in the basement (where it was originally discovered), pending proper disposal. The on-contact reading was 4 mR/hr."


Power Reactor
Event Number: 55311
Facility: Palisades
Region: 3     State: MI
Unit: [1] [] []
RX Type: [1] CE
NRC Notified By: Barb Dotson
HQ OPS Officer: Thomas Kendzia
Notification Date: 06/16/2021
Notification Time: 23:41 [ET]
Event Date: 06/16/2021
Event Time: 15:50 [EDT]
Last Update Date: 06/17/2021
Emergency Class: Non Emergency
10 CFR Section:
50.72(b)(3)(v)(D) - Accident Mitigation
Person (Organization):
HILLS, DAVID (R3)
Power Reactor Unit Info
Unit SCRAM Code RX Crit Initial PWR Initial RX Mode Current PWR Current RX Mode
1 N Y 100 Power Operation 100 Power Operation
Event Text
EN Revision Imported Date: 7/16/2021

EN Revision Text: ATMOSPHERIC STEAM DUMP VALVES INOPERABLE

"On June 16, 2021, at 1550 EDT, Palisades Nuclear Plant was operating in Mode 1 at 100% power. At that time, operations identified an acrid odor in the control room. Investigation revealed that the steam dump control relay had failed, rendering all four atmospheric steam dump valves inoperable.

"The loss of function of all four atmospheric steam dump valves is reportable in accordance with 10 CFR 50.72(b)(3)(v)(D) as an event or condition that at the time of discovery could have prevented the fulfillment of the safety function of structures or systems that are needed to mitigate the consequences of an accident.

"Troubleshooting and replacement of the relay are in progress.

"The plant remains stable in Mode 1 at 100% power.

"The NRC Resident Inspector has been notified."

Unit 1 is in a 24 hour LCO for Tech Spec 3.7.4.b, atmospheric steam dump valve inoperability. The Unit is in a normal offsite power line-up.


Agreement State
Event Number: 55327
Rep Org: TEXAS DEPT OF STATE HEALTH SERVICES
Licensee: Oceaneering International, Inc.
Region: 4
City: Houston   State: TX
County:
License #: L 06845
Agreement: Y
Docket:
NRC Notified By: Arthur Tucker
HQ OPS Officer: Donald Norwood
Notification Date: 06/25/2021
Notification Time: 13:47 [ET]
Event Date: 06/16/2021
Event Time: 00:00 [CDT]
Last Update Date: 06/25/2021
Emergency Class: Non Emergency
10 CFR Section:
Agreement State
Person (Organization):
KELLAR, RAY (R4)
NMSS_EVENTS_NOTIFICATION, (EMAIL)
Event Text
EN Revision Imported Date: 7/23/2021

EN Revision Text: AGREEMENT STATE REPORT - RADIOGRAPHY SOURCE DISCONNECT

The following information was received via E-mail:

"On June 25, 2021, the Agency [Texas Department of State Health Services] was notified by the licensee's Radiation Safety Officer (RSO) that while conducting radiography in their shooting bay, they experienced a source disconnect. The disconnect involved a QSA 880D exposure device containing a 60 curie iridium-192 source.

"The RSO stated the radiographer had completed an exposure and was entering the bay to exchange the film. As they passed the entrance beam the radiation alarm went off. The radiographer exited the area. The licensee was unable to retract the source. They contacted a service company who came to the licensee's location.

"It was determined that the ball on the drive cable had broken free of the drive cable. The service company was able to retract the source into the exposure device. No overexposures occurred due to this event. Additional information will be provided as it is received in accordance with SA-300."

Texas Incident Number: I-9860.


Agreement State
Event Number: 55472
Rep Org: COLORADO DEPT OF HEALTH
Licensee: Davis Partnership Architects PC
Region: 4
City: Denver   State: CO
County:
License #: GL000674
Agreement: Y
Docket:
NRC Notified By: Kathryn Mote
HQ OPS Officer: Donald Norwood
Notification Date: 09/17/2021
Notification Time: 13:46 [ET]
Event Date: 06/16/2021
Event Time: 00:00 [MDT]
Last Update Date: 09/17/2021
Emergency Class: Non Emergency
10 CFR Section:
Agreement State
Person (Organization):
AZUA, RAY (R4)
NMSS_EVENTS_NOTIFICATION, (EMAIL)
ILTAB, (EMAIL)
Event Text
EN Revision Imported Date: 10/15/2021

EN Revision Text: AGREEMENT STATE REPORT - LOST TRITIUM EXIT SIGN

The following was received from the Colorado Department of Health via email:

"One exit sign, containing 7.5 Ci of tritium, was reported as lost during annual registration."

NMED No.: CO210028

THIS MATERIAL EVENT CONTAINS A 'Less than Cat 3' LEVEL OF RADIOACTIVE MATERIAL

Sources that are "Less than IAEA Category 3 sources," are either sources that are very unlikely to cause permanent injury to individuals or contain a very small amount of radioactive material that would not cause any permanent injury. Some of these sources, such as moisture density gauges or thickness gauges that are Category 4, the amount of unshielded radioactive material, if not safely managed or securely protected, could possibly - although it is unlikely - temporarily injure someone who handled it or were otherwise in contact with it, or who were close to it for a period of many weeks. For additional information go to http://www-pub.iaea.org/MTCD/publications/PDF/Pub1227_web.pdf


Agreement State
Event Number: 55465
Rep Org: ARKANSAS DEPARTMENT OF HEALTH
Licensee: 3D Imaging Drug Design Development
Region: 4
City: Little Rock   State: AR
County:
License #: ARK-1008-03214
Agreement: Y
Docket:
NRC Notified By: Chris Talley
HQ OPS Officer: Brian Lin
Notification Date: 09/15/2021
Notification Time: 15:17 [ET]
Event Date: 06/16/2021
Event Time: 00:00 [CDT]
Last Update Date: 09/15/2021
Emergency Class: Non Emergency
10 CFR Section:
Agreement State
Person (Organization):
AZUA, RAY (R4)
NMSS_EVENTS_NOTIFICATION, (EMAIL)
ILTAB, (EMAIL)
Event Text
EN Revision Imported Date: 10/15/2021

EN Revision Text: AGREEMENT STATE REPORT - MISSING RADIOACTIVE SHIPMENT

The following information was received from the state of Arkansas (the Department) via email:

"The Department received notification on June 16, 2021, from licensee 3D Imaging Drug Design and Development, 3DI, that a package containing 87 milliCi of Zr-89 did not reach its shipping destination.

"Upon review of the event, the Licensee transferred the package via [common carrier] pick-up from its Little Rock, Arkansas campus on Saturday June 12, 2021, with a scheduled delivery date of June 14, 2021. The package destination was:

"Memorial Sloan Kettering
1275 York Avenue
New York City, New York

"On June 16, 2021, the package was reported missing to the Department and the activity of the package was estimated to be approximately 30 milliCi based on the 3.2 day half-life of Zr-89.

"The licensee had contacted the [common carrier] and the [common carrier] stated that they were unable to locate the package. Department inspectors visited the 3DI campus on June 18, 2021 and questioned the Licensee about the lost package. At that time, and currently, the [common carrier] has been unable to locate the package. The Department has spoken with multiple individuals associated with the [common carrier], both at the Little Rock location and with the companies Dangerous Goods Administration Department, about the lost package. The Department has also followed up with 3DI.

"The associates at the [common carrier] were able to confirm that the package was received and scanned. After acceptance, and the initial scan of the package, there are no other scans available to determine what has happened to the package. 3DI stated that the company has experienced additional shipping problems since the lost package. 3DI has had additional packages that have been scanned and shipped to Licensees other than the intended recipient. All additional packages have been able to be located and recovered or reach their final destination successfully.

"The initial package of Zr-89 remains missing. The Department has requested and is awaiting the 30 Day report from 3DI as required. The Department considers this incident open pending location of the missing package and will update this report as information is received. "

AR incident no.: AR-2021-007

THIS MATERIAL EVENT CONTAINS A 'Less than Cat 3' LEVEL OF RADIOACTIVE MATERIAL

Sources that are "Less than IAEA Category 3 sources," are either sources that are very unlikely to cause permanent injury to individuals or contain a very small amount of radioactive material that would not cause any permanent injury. Some of these sources, such as moisture density gauges or thickness gauges that are Category 4, the amount of unshielded radioactive material, if not safely managed or securely protected, could possibly - although it is unlikely - temporarily injure someone who handled it or were otherwise in contact with it, or who were close to it for a period of many weeks. For additional information go to http://www-pub.iaea.org/MTCD/publications/PDF/Pub1227_web.pdf"


Agreement State
Event Number: 55312
Rep Org: CALIFORNIA RADIATION CONTROL PRGM
Licensee: UCI Irvine Medical Center
Region: 4
City: Orange   State: CA
County:
License #: 0278-30
Agreement: Y
Docket:
NRC Notified By: Donald Oesterle
HQ OPS Officer: Thomas Kendzia
Notification Date: 06/17/2021
Notification Time: 10:49 [ET]
Event Date: 06/15/2021
Event Time: 00:00 [PDT]
Last Update Date: 06/17/2021
Emergency Class: Non Emergency
10 CFR Section:
Agreement State
Person (Organization):
JOSEY, JEFFREY (R4)
NMSS_EVENTS_NOTIFICATION, (EMAIL)
Event Text
EN Revision Imported Date: 7/16/2021

EN Revision Text: AGREEMENT STATE REPORT - PORTION OF THE DOSE DELIVERED TO THE WRONG LOCATION

The following was received from the California Department of Public Health via email:

"On June 16, 2021, the Radiologic Health Branch was notified of a reportable medical event that occurred on June 15, 2021 during a patients' liver metastases treatment with Y-90 Sirtex SIRSpheres. The AU's treatment plan called for treating the left lobe of the liver with 0.29-0.83 GBq of Y-90 SIRSpheres. The reason for the range was that if the liver became saturated, the treatment would be stopped at that point. During the treatment, periodic flushing cycles with contrast and flouroscopy were performed. At a mid-way point, the team discovered contrast material in the right liver lobe, indicating the microcather had moved from the left artery to the right artery. Upon discovery, the procedure was stopped, the microcatherter was removed and a new one was placed and the they began to infuse the left liver lobe again with the remaining Y-90 SIRSpheres without incident. Post treatment, a bremsstrahlung image of the two liver lobes indicated that both lobes had received Y-90 activity. The Radiation Oncologist estimated that the left lobe received less than the intended Y-90 activity. The right lobe received between 33%-67% of the Y-90 activity.

"The actual dose to either lobe has not been calculated, but based on dosimetry information in the package insert, the dose to the right lobe was > 0.5 Gy ( 50 rem). Treatment of the liver's right lobe was not intended during this procedure. On June 1, 2021, the patient had been treated with Y-90 SIRSpheres of the patient's right liver lobe. The patient was informed of the issue and there are no negative consequences expected for the patient."

California Item Number: 061621

A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.



Agreement State
Event Number: 55315
Rep Org: NJ DEPT OF ENVIRONMENTAL PROTECTION
Licensee: Private Citizen
Region: 1
City: Mountain Lakes   State: NJ
County:
License #: N/A
Agreement: Y
Docket:
NRC Notified By: Richard Peros
HQ OPS Officer: Jeffrey Whited
Notification Date: 06/17/2021
Notification Time: 18:07 [ET]
Event Date: 06/15/2021
Event Time: 00:00 [EDT]
Last Update Date: 06/17/2021
Emergency Class: Non Emergency
10 CFR Section:
Agreement State
Person (Organization):
LILLIENDAHL, JON (R1)
NMSS_EVENTS_NOTIFICATION, (EMAIL)
Event Text
EN Revision Imported Date: 7/16/2021

EN Revision Text: AGREEMENT STATE REPORT - PRIVATE CITIZEN IN POSSESSION OF NUCLEAR MATERIAL

The following was received from the New Jersey Department of Environmental Protection (DEP) via email:

"On Tuesday evening, June 15, 2021, the DEP was notified that a member of the public had come to the Mountain Lakes police HQ to report that he had in his possession some radioactive material that he had obtained 60 years ago when he worked for Westinghouse as an engineer. The material was reported to be "Nuclear Reactor grade U-238 with 5 percent U-235 enriched". The citizen further stated that the material is wrapped in lead foil and placed in a lead pipe with the ends pinched over. The material allegedly consists of three or four rejected pellets, approximately 3/8 inch diameter. These were rejected because of dimensional irregularities. They are allegedly doughnut shaped with a hole in the center. They were reportedly to be used in Westinghouse nuclear reactors/steam power. The citizen stated that the material in question has been stored for decades in a lead pipe, sealed off at the ends, and then tightly wrapped in lead sheeting. It was also clearly labelled with the word "Radioactive" and then placed inside a large can, which has been securely stored in the citizen's garage for several decades. DEP personnel responded to the citizen's home on June 17, 2021. The material was found stored as the citizen had previously described. The container was not opened. There was no detectable removable contamination on the outside of the container. The material was returned to the garage where it will be secured pending proper disposal."


Power Reactor
Event Number: 55306
Facility: Fort Calhoun
Region: 4     State: NE
Unit: [1] [] []
RX Type: (1) CE
NRC Notified By: Bill Rice
HQ OPS Officer: Jeffrey Whited
Notification Date: 06/15/2021
Notification Time: 15:38 [ET]
Event Date: 06/15/2021
Event Time: 12:30 [CDT]
Last Update Date: 06/15/2021
Emergency Class: Non Emergency
10 CFR Section:
50.72(b)(2)(xi) - Offsite Notification
Person (Organization):
JOSEY, JEFFREY (R4)
Power Reactor Unit Info
Unit SCRAM Code RX Crit Initial PWR Initial RX Mode Current PWR Current RX Mode
1 N N 0 Defueled 0 Defueled
Event Text
EN Revision Imported Date: 7/15/2021

EN Revision Text: OFFSITE NOTIFICATION

"At 1230 CDT a report was made to the State of Nebraska Department of Environment and Energy (NDEE) based on the analytical report for soil samples from the area surrounding the removed FO-1, Emergency Diesel Generator Fuel Oil Storage Tank, and the removed FO-32, TSC/Security Fuel Oil Tank. The tanks were removed as part of Fort Calhoun Station decommissioning and soil samples were tested due to soil discoloration at the time the tanks were pulled. The soil contamination levels are from the historic use of the tank. The contamination levels are above the lab reporting limits and thereby reportable to the State of Nebraska Department of Environment and Energy. The NDEE will determine what, if any, remediation may be required. The state NDEE requested the District utilize their Spill Form because this is the simplest method of State notification for tanks exempted due to 40CFR280.10(c)(4).

"No active petroleum spills are in progress and appropriate remediation actions will be taken in accordance with Nebraska State regulation and guidance."

The licensee notified the NRC Region IV Office.


Non-Agreement State
Event Number: 55308
Rep Org: G2 Consulting
Licensee: G2 Consulting
Region: 3
City: Troy   State: MI
County:
License #: 21-26593-01
Agreement: N
Docket:
NRC Notified By: Bruce Wibeding
HQ OPS Officer: Thomas Herrity
Notification Date: 06/15/2021
Notification Time: 15:57 [ET]
Event Date: 06/15/2021
Event Time: 15:30 [EDT]
Last Update Date: 06/15/2021
Emergency Class: Non Emergency
10 CFR Section:
20.2201(a)(1)(i) - Lost/Stolen Lnm>1000x
Person (Organization):
HILLS, DAVID (R3)
NMSS_EVENTS_NOTIFICATION, (EMAIL)
ILTAB, (EMAIL)
CNSC (CANADA), - (EMAIL)
Event Text
EN Revision Imported Date: 7/15/2021

EN Revision Text: NON-AGREEMENT STATE - LOST MOISTURE DENSITY GAUGE

The following was received from the licensee via email:

"On Monday afternoon June 14, 2021, [the licensee] went to perform a leak test on Troxler gauge 70894 (Model No. 3430; Cs-137 8 milliCurie; Am-241/Be 40 milliCurie), which was due for a leak test the same date 6/14/21. [The licensee] looked in [their] system to see which employee had the gauge. It was not checked out by anyone that day so [the licensee] went through the storage room in Ann Arbor and then Troy to find the gauge and perform the test. The gauge was not present in either location. [The licensee] began making phone calls to employees to see if an employee had the gauge and it was just not checked out properly in [their] system.

"[The licensee] reviewed [their] system and determined the gauge was last used by a former employee on January 5, 2021. [The licensee] was able to reach that person and he indicated he returned the gauge to the locked storage room in [the] Ann Arbor office on January 6, 2021. The gauge was not shown as having been used by any employee since that date within [the] system. [The licensee does] not know when the gauge went missing or how the gauge was removed from the office. [The licensee] also cannot confirm if the former employee actually returned the gauge on January 6, 2021. Calibration and leak tests were performed in December 2020 by ATSNUC, Inc. [The licensee] contacted all of the service facilities that [they] use to see if it might be in for service but [the licensee has] been unable to locate the gauge. [The licensee is] also checking with all service centers listed on the APNGA website to see if by chance some other party has sent in/returned this gauge to one of those facilities. It was determined today, 6/15/21, at approximately 1534 EDT (less than 24 hours after the initial discovery) that [the licensee] could not locate the gauge."

The licensee notified the NRC Region III Office.

THIS MATERIAL EVENT CONTAINS A 'Less than Cat 3' LEVEL OF RADIOACTIVE MATERIAL

Sources that are "Less than IAEA Category 3 sources," are either sources that are very unlikely to cause permanent injury to individuals or contain a very small amount of radioactive material that would not cause any permanent injury. Some of these sources, such as moisture density gauges or thickness gauges that are Category 4, the amount of unshielded radioactive material, if not safely managed or securely protected, could possibly - although it is unlikely - temporarily injure someone who handled it or were otherwise in contact with it, or who were close to it for a period of many weeks. For additional information go to http://www-pub.iaea.org/MTCD/publications/PDF/Pub1227_web.pdf


Agreement State
Event Number: 55310
Rep Org: ILLINOIS EMERGENCY MGMT. AGENCY
Licensee: University of Chicago Hospital
Region: 3
City: Chicago   State: IL
County:
License #: IL-01678-02
Agreement: Y
Docket:
NRC Notified By: Gary Forsee
HQ OPS Officer: Jeffrey Whited
Notification Date: 06/16/2021
Notification Time: 13:10 [ET]
Event Date: 06/15/2021
Event Time: 00:00 [CDT]
Last Update Date: 06/16/2021
Emergency Class: Non Emergency
10 CFR Section:
Agreement State
Person (Organization):
HILLS, DAVID (R3)
NMSS_EVENTS_NOTIFICATION, (EMAIL)
Event Text
EN Revision Imported Date: 7/16/2021

EN Revision Text: AGREEMENT STATE REPORT - MEDICAL EVENT - UNDERDOSE

The following was received from the Illinois Emergency Management Agency (the Agency) via email:

"University of Chicago Medical Center contacted the Agency on the afternoon of 6/15/21 to report a medical underdose of Lu-177 that occurred that day. Although information provided was preliminary, no untoward medical impact is expected to the patient. The Radiation Safety Officer (RSO) for the licensee contacted the Agency at approximately 1615 CDT on June 15, 2021, to report that a patient scheduled to receive 200 mCi of Lu-177, Lutathera therapy for neuroendocrine tumors, received only 68 percent of the dose prescribed (136 mCi) in the written directive. The underdosing was reportedly due to leakage in the adaptor/needle connection. No personnel or area contamination occurred. The licensee is still evaluating whether or not the remaining dose will be delivered at a future date. The RSO confirmed the patient and referring physician were notified within 24 hours.

"IEMA inspectors will perform a reactive inspection on June 17, 2021. The reporting requirements for the licensee, as specified in 32 Ill. Adm. Code 335.1080(c) were met, and the licensee is aware of the need for a written report within 15 days. This report will be updated once additional details become available on 6/17/21."

Item Number: IL210019

A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.