Due to a lapse in appropriations, the NRC has ceased normal operations. However, excepted and exempted activities necessary to maintain critical health and safety functions—as well as essential progress on designated critical activities, including those specified in Executive Order 14300—will continue, consistent with the OMB-Approved NRC Lapse Plan.

Event Notification Report for March 21, 2019

U.S. Nuclear Regulatory Commission
Operations Center

Event Reports For
3/20/2019 - 3/21/2019

** EVENT NUMBERS **


53931


Agreement State Event Number: 53931
Rep Org: OHIO BUREAU OF RADIATION PROTECTION
Licensee: CARDINAL HEALTH
Region: 3
City: DUBLIN   State: OH
County:
License #: 02500250000
Agreement: Y
Docket:
NRC Notified By: MICHAEL RUBADUE
HQ OPS Officer: MARK ABRAMOVITZ
Notification Date: 03/13/2019
Notification Time: 08:46 [ET]
Event Date: 02/26/2019
Event Time: 00:00 [EDT]
Last Update Date: 03/13/2019
Emergency Class: NON EMERGENCY
10 CFR Section:
AGREEMENT STATE
Person (Organization):
DARIUSZ SZWARC (R3DO)
NMSS_EVENTS_NOTIFICATION (EMAIL)

Event Text

AGREEMENT STATE REPORT - Mo-99/Tc-99 BREAKTHROUGH LIMIT EXCEEDED.

The following report was received via e-mail:

"On March 7, 2019, the Ohio Department of Health received a phone notification from Cardinal Health stating the first elution on 2/26/2019 from the NorthStar RadioGenix Mo-99/Tc-99m generator exceeded breakthrough limits in Ohio Administrative Code 3701:1-58-35 (10 CFR 35.204). The measured breakthrough was 0.252 microCuries Mo-99 per milliCurie Tc-99m. Cardinal Health stated this was an isolated event and all other elutions have been within the limits. The licensee is using the elution for research only; it is not for human use. Cardinal Health has contacted the manufacture to determine the cause."

Ohio Event: 190004