Event Notification Report for July 03, 2006

U.S. Nuclear Regulatory Commission
Operations Center

EVENT REPORTS FOR
07/02/2006 - 07/03/2006

EVENT NUMBERS
42707

General Information or Other
Event Number: 42707
Rep Org: MARYLAND DEPT OF THE ENVIRONMENT
Licensee: ANNE ARUNDEL COUNTY MEDICAL CENTER
Region: 1
City: ANNAPOLIS   State: MD
County: ANNE ARUNDEL
License #: MD-0300106
Agreement: Y
Docket:
NRC Notified By: RAY MANLEY
HQ OPS Officer: PETE SNYDER
Notification Date: 07/17/2006
Notification Time: 14:55 [ET]
Event Date: 07/03/2006
Event Time: 00:00 [EDT]
Last Update Date: 07/18/2006
Emergency Class: NON EMERGENCY
10 CFR Section:
Person (Organization):
JOHN KINNEMAN (R1)
GREG MORELL (NMSS)
Event Text
AGREEMENT STATE REPORT - MEDICAL MISADMINISTRATION USING HIGH DOSE RATE AFTERLOADER

On 7/17/06 the State of Maryland received a report from the Anne Arundel Medical Center in Annapolis, Maryland of a potential medical misadministration using a high dose rate afterloader.

Approximately two to three weeks ago a patient was administered a dose using a Varian VariSource 200 HDR brachytherapy afterloader to treat a lung tumor. During the dose administration the catheter was determined to be about ten centimeters short of the planned location for the dose. Because of the mispositioning or the catheter, an unintended dose of less than 100 centigray was given to patient's vocal chord area. The lungs received a dose of 5 gray. The licensee notified the patient of the issue. A licensee physician examined the patient today and determined that there was no erythema and that there would be no adverse medical effects in the unintended dose area. The licensee reported that a human error not a device error resulted in the misadministration.

The State will be conducting follow-up investigations into this incident to include reviewing the licensee's determination of the cause, as well as, determining why the event was not reported until 2 to 3 weeks after it had occurred.

Typically a 10 curie Iridium-192 source is used in this type of instrument.

* * * UPDATE AT 1100 EDT ON 07/18/06 FROM RAY MANLEY TO S. SANDIN * * *

The patient did not receive a dose of 5 gray to the lungs as described above. This was the prescribed dose. Also, the information is preliminary pending completion of the follow-up investigation. Notified R1DO (Kinneman) and NMSS (Morell).