Due to a lapse in appropriations, the NRC has ceased normal operations. However, excepted and exempted activities necessary to maintain critical health and safety functions—as well as essential progress on designated critical activities, including those specified in Executive Order 14300—will continue, consistent with the OMB-Approved NRC Lapse Plan.

Event Notification Report for February 20, 2002

                         *** NOT FOR PUBLIC DISTRIBUTION ***
                    U.S. Nuclear Regulatory Commission
                              Operations Center

                              Event Reports For
                           02/19/2002 - 02/20/2002

                              ** EVENT NUMBERS **

38707  38708  38709  

+------------------------------------------------------------------------------+
|Power Reactor                                    |Event Number:   38707       |
+------------------------------------------------------------------------------+
                         *** NOT FOR PUBLIC DISTRIBUTION ***
+------------------------------------------------------------------------------+
+------------------------------------------------------------------------------+
| FACILITY: BEAVER VALLEY            REGION:  1  |NOTIFICATION DATE: 02/19/2002|
|    UNIT:  [] [2] []                 STATE:  PA |NOTIFICATION TIME: 01:27[EST]|
|   RXTYPE: [1] W-3-LP,[2] W-3-LP                |EVENT DATE:        02/18/2002|
+------------------------------------------------+EVENT TIME:        18:15[EST]|
| NRC NOTIFIED BY:  JAMES WITTER                 |LAST UPDATE DATE:  02/19/2002|
|  HQ OPS OFFICER:  FANGIE JONES                 +-----------------------------+
+------------------------------------------------+PERSON          ORGANIZATION |
|EMERGENCY CLASS:          NON EMERGENCY         |JAMES TRAPP          R1      |
|10 CFR SECTION:                                 |                             |
|AINA 50.72(b)(3)(v)(A)   POT UNABLE TO SAFE SD  |                             |
|AINB 50.72(b)(3)(v)(B)   POT RHR INOP           |                             |
|                                                |                             |
|                                                |                             |
+-----+----------+-------+--------+-----------------+--------+-----------------+
|UNIT |SCRAM CODE|RX CRIT|INIT PWR|   INIT RX MODE  |CURR PWR|  CURR RX MODE   |
+-----+----------+-------+--------+-----------------+--------+-----------------+
|                                                   |                          |
|2     N          N       0        Cold Shutdown    |0        Cold Shutdown    |
|                                                   |                          |
+------------------------------------------------------------------------------+
                                   EVENT TEXT                                   
+------------------------------------------------------------------------------+
| BOTH EMERGENCY DIESEL GENERATORS INADVERTENTLY MADE INOPERABLE               |
|                                                                              |
| "On 02/18/02, at 1815 hours, the licensee discovered that the Beaver Valley  |
| Unit 2 #1 Emergency Diesel Generator had been rendered inoperable            |
| inadvertently, by the posting of a clearance at 1408 hours on 2/14/02. The   |
| #2 Emergency Diesel Generator had previously been removed from service for   |
| maintenance activities. Upon discovery that the #1 Emergency Diesel          |
| Generator was inoperable, action was immediately initiated to restore the #1 |
| Emergency Diesel Generator to Operable status. These actions were completed  |
| and the #1 Emergency Diesel Generator was restored to Operable status at     |
| 1830 hrs on 02/18/02.                                                        |
|                                                                              |
| "The #1 Diesel Generator was rendered inoperable at 1406 hrs on 02/14/02.    |
| During the period of inoperability, the licensee conducted core alterations  |
| including core reload during the ongoing ninth refueling outage. During the  |
| affected period, there were no challenges to the offsite power sources, and  |
| core cooling was maintained. Investigation into this event is ongoing.       |
|                                                                              |
| "This event is being reported under 10CFR50.72(v)(A) and 10CFR50.72(v)(B) as |
| a loss of Safe Shutdown Capability, and RHR Capability, respectively. There  |
| were no radiological releases associated with this event,"                   |
|                                                                              |
| The licensee intends to notify the NRC Resident Inspector.                   |
+------------------------------------------------------------------------------+

!!!!!!!!! THIS EVENT HAS BEEN RETRACTED. THIS EVENT HAS BEEN RETRACTED  !!!!!!!
+------------------------------------------------------------------------------+
|Hospital                                         |Event Number:   38708       |
+------------------------------------------------------------------------------+
                         *** NOT FOR PUBLIC DISTRIBUTION ***
+------------------------------------------------------------------------------+
+------------------------------------------------------------------------------+
| REP ORG:  UNIVERSITY OF WISCONSIN AT MADISON   |NOTIFICATION DATE: 02/19/2002|
|LICENSEE:  UNIVERSITY OF WISCONSIN AT MADISON   |NOTIFICATION TIME: 08:12[EST]|
|    CITY:  MADISON                  REGION:  3  |EVENT DATE:        02/15/2002|
|  COUNTY:                            STATE:  WI |EVENT TIME:             [CST]|
|LICENSE#:  4809843-18            AGREEMENT:  N  |LAST UPDATE DATE:  02/19/2002|
|  DOCKET:                                       |+----------------------------+
|                                                |PERSON          ORGANIZATION |
|                                                |ANTON VEGEL          R3      |
|                                                |JOHN HICKEY          NMSS    |
+------------------------------------------------+                             |
| NRC NOTIFIED BY:  BEN-ZIKRI                    |                             |
|  HQ OPS OFFICER:  CHAUNCEY GOULD               |                             |
+------------------------------------------------+                             |
|EMERGENCY CLASS:          NON EMERGENCY         |                             |
|10 CFR SECTION:                                 |                             |
|LADM 35.33(a)            MED MISADMINISTRATION  |                             |
|                                                |                             |
|                                                |                             |
|                                                |                             |
|                                                |                             |
+------------------------------------------------------------------------------+

                                   EVENT TEXT                                   
+------------------------------------------------------------------------------+
| MEDICAL MISADMINISTRATION DUE TO AN EQUIPMENT MALFUNCTION                    |
|                                                                              |
| A BetaCath Systems device containing strontium-90 produced by Novoste was    |
| being used to treat a patient.  On sending the active sources to the target  |
| area, a fluid leak occurred which slowed the movement of the seeds into the  |
| target vessel/position (Left Anterior Descending Artery). Not able to see    |
| the active seeds on the fluoroscopy monitors, a decision to have the active  |
| seeds returned to the holding device was made.   A total time of 17 seconds  |
| elapsed from the time of delivery until the active seeds were returned to    |
| the holding device.                                                          |
|                                                                              |
| On close inspection of the treatment catheter/device, one of the sealing     |
| O-rings was found to be ruptured.  This was the ultimate cause of the fluid  |
| leak and consequently the slow movement of the active seeds to the target    |
| position.                                                                    |
|                                                                              |
| With a dose rate of 0.095 Gy/s at 2 mm, the artery wall along the            |
| approximately 1 m track over which the source traveled out and back for its  |
| 17 sec would be calculated as:                                               |
|                                                                              |
| Speed equals 2 * (100 cm)/17 s = 11.8 cm/s                                   |
|                                                                              |
| Time wall was exposed to the 4 cm source = 4 cm / 11.8 cm/s = 0.34 s, but    |
| this time is doubled since it is exposed coming and going, to 0.68 s.        |
|                                                                              |
| Dose to the wall = 0.68 s * 0.095 Gy/s = 0.0646 Gy = 6.5 rads.               |
|                                                                              |
| The normal dose to the wall (assuming an 8 s round trip) would be 3.0 rads.  |
|                                                                              |
| A new catheter was tested with the non-active device and the treatment was   |
| continued.  The treatment was timed, checked for the return of the active    |
| sources and surveyed the patient/room post treatment.   No problems were     |
| encountered with the new treatment catheter.                                 |
|                                                                              |
| * * * UPDATE 1745EST ON 2/19/02 FROM ABDUL BEN-ZIKRI TO S. SANDIN VIA E-MAIL |
| * * *                                                                        |
|                                                                              |
| The following is a portion of an e-mail received at hoo1@nrc.gov:            |
|                                                                              |
| "I am retracting my event report (Event # 38708).  I took a conservative     |
| action and reported this event based on preliminary written information.     |
|                                                                              |
| "Our Hospital staff followed all standard operating & emergency procedures,  |
| and retracted the source when they realized that they did not arrive to the  |
| area of clinical interest as indicated in  the report below.  Based on the   |
| investigation analysis, we conclude that there was no misadministration to   |
| report."                                                                     |
|                                                                              |
| Notified R3DO(Jorgensen) and NMSS(Brach).                                    |
| .                                                                            |
+------------------------------------------------------------------------------+

+------------------------------------------------------------------------------+
|Other Nuclear Material                           |Event Number:   38709       |
+------------------------------------------------------------------------------+
                         *** NOT FOR PUBLIC DISTRIBUTION ***
+------------------------------------------------------------------------------+
+------------------------------------------------------------------------------+
| REP ORG:  PRINCETON UNIVERSITY                 |NOTIFICATION DATE: 02/19/2002|
|LICENSEE:  PRINCETON UNIVERSITY                 |NOTIFICATION TIME: 11:35[EST]|
|    CITY:  PRINCETON                REGION:  1  |EVENT DATE:        01/23/2002|
|  COUNTY:                            STATE:  NJ |EVENT TIME:             [EST]|
|LICENSE#:  290518524             AGREEMENT:  N  |LAST UPDATE DATE:  02/19/2002|
|  DOCKET:                                       |+----------------------------+
|                                                |PERSON          ORGANIZATION |
|                                                |MICHAEL MODES        R1      |
|                                                |JOHN HICKEY          NMSS    |
+------------------------------------------------+                             |
| NRC NOTIFIED BY:  SUE DUPRE                    |                             |
|  HQ OPS OFFICER:  FANGIE JONES                 |                             |
+------------------------------------------------+                             |
|EMERGENCY CLASS:          NON EMERGENCY         |                             |
|10 CFR SECTION:                                 |                             |
|BAB2 20.2201(a)(1)(ii)   LOST/STOLEN LNM>10X    |                             |
|                                                |                             |
|                                                |                             |
|                                                |                             |
|                                                |                             |
+------------------------------------------------------------------------------+

                                   EVENT TEXT                                   
+------------------------------------------------------------------------------+
| MISSING RAT CARCASSES CONTAMINATED WITH TRITIUM                              |
|                                                                              |
| On 1/23/02 the Radiation Safety Officer was informed that 15 rat carcasses   |
| contaminated with 14 mCi of tritium were missing from the freezer.  This was |
| discovered on 1/21/02 when a scheduled pickup for disposal was made and the  |
| 2 bags of rat carcasses were not located.  A thorough search of the          |
| laboratories and freezers was made.  It is believed that the 2 bags were     |
| mistakenly picked up on 11/7/01 when a scheduled pickup for disposal of      |
| non-contaminated rat carcasses was made.  The 2 bags were last seen in       |
| October when the bags were placed in the freezer.                            |
|                                                                              |
| New procedures have been put in place to prevent this event from being       |
| repeated.  A written report will be submitted within 30 days.                |
+------------------------------------------------------------------------------+