Part 21 Report - 2000-260
REP ORG: WESTRONICS |
|
NOTIFICATION DATE: |
11/06/2000 |
LICENSEE: WESTRONICS |
|
NOTIFICATION TIME: |
14:05[EST] |
CITY: KINGWOOD |
REGION: 4 |
EVENT DATE: |
11/03/2000 |
COUNTY: |
STATE: TX |
EVENT TIME: |
12:00[CST] |
LICENSE#: |
AGREEMENT: Y |
LAST UPDATE DATE: |
11/06/2000 |
DOCKET: |
|
PERSON |
ORGANIZATION |
|
|
JOE TAPIA |
R4 |
|
|
BRIAN BONSER |
R2 |
|
|
VERN HODGE |
NRR |
NRC NOTIFIED BY: |
ROBERT AGEE |
|
|
HQ OPS OFFICER: |
JOHN MacKINNON |
|
|
EMERGENCY CLASS: |
N/A |
|
|
10 CFR SECTION: |
|
|
|
CCCC 21.21 |
UNSPECIFIED PARAGRAPH |
|
|
|
|
|
|
EVENT TEXT
WESTRONICS CSVC VIDEOGRAPHIC RECORDERS
On or about November 3, 2000, Westronics determined that Westronics Series CSVC "Smartview" Nuclear Safety Related videographic recorders, if utilized to measure thermocouple inputs, require hardware modification to ensure accurate measurement. The hardware modification requires installation of capacitors associated with the Dallas 1620 Temperature Compensation Chips. Instruments which would require the modification were shipped between July 16, 1998 and July 7, 2000. Nuclear Power Plants affected are Riverbend, Shearon Harris, and Brunswick.
Westronics Corrective Action includes installation of the capacitors.
Affected Customers will be notified in accordance with the requirements of 10CFR21.
Page Last Reviewed/Updated Wednesday, March 24, 2021