Part 21 Report - 1998-670
LICENSEE: |
ROSEMOUNT NUCLEAR INSTRUMENTS, INC |
SITE: |
ROSEMOUNT NUCLEAR INSTRUMENTS, |
EN NUMBER: |
34923 |
DOCKET: |
|
|
EVENT DATE: |
10-16-98 |
RX TYPE: |
|
|
EVENT TIME: |
00:00 |
VENDORS: |
|
|
NOTIFY DATE: |
10-16-98 |
EMERGENCY CLASS: N/A |
REGION: 3 |
STATE: MN |
TIME: |
14:40 |
OPS OFFICER: |
BOB STRANSKY |
|
|
|
10 CFR SECTION: |
CCCC 21.21 |
UNSPECIFIED PARAGRAPH |
10 CFR PART 21 REPORT REGARDING INCORRECTLY TABULATED INFORMATION REGARDING RADIATION PERFORMANCE.
"PURSUANT TO 10CFR PART 21 PARAGRAPH 21.21(b), ROSEMOUNT NUCLEAR INSTRUMENTS, INC. (RNII) IS WRITING TO INFORM YOU THAT CERTAIN DATA CONTAINED IN THE QUALIFICATION AND TYPE TEST REPORTS FOR ROSEMOUNT MODEL 1154 SERIES H ALPHALINE NUCLEAR PRESSURE
TRANSMITTER, WHICH WERE USED TO SPECIFY THE RADIATION PERFORMANCE IDENTIFIED IN RELEVANT PRODUCT DATA SHEETS AND MANUALS, WERE INCORRECTLY TABULATED AND REPORTED. THIS RESULTS IN A SMALL CHANGE TO THE 1154 SERIES H NUCLEAR PRESSURE TRANSMITTER 30-MINUTE RADIATION ACCURACY SPECIFICATION. ALL OTHER RADIATION AND OTHER ACCIDENT LEVEL SPECIFICATIONS ARE UNAFFECTED."
Page Last Reviewed/Updated Wednesday, March 24, 2021