U.S. Nuclear Regulatory Commission
Relaxation of Definition of Source Check in Reference to Effluent Radiation Monitors
HPPOS-229 PDR-9111210328
Title: Relaxation of Definition of Source Check in
Reference to Effluent Radiation Monitors
See the memorandum from L. J. Cunningham to J. H. Joyner
(and others) dated December 6, 1990. This memo states that
any proposal by a licensee to relax the definition of a
source check is not acceptable without compensatory
measures to maintain overall effluent control for the
proposed relaxation.
A licensee had submitted an amendment request to move the
existing procedural details of the current Radiological
Effluent Technical Specifications (RETS) to the Offsite
Dose Calculation Manual (ODCM). The licensee, as well as
twenty-two other facilities, used plastic
scintillator / photomultiplier type effluent radiation
monitors that contained either a built-in LED light source
or a secondary check source that did not expose the primary
detector. These alternative source check measurements were
used to meet the monthly qualitative source check
requirement. The definition of "source check" under the
Technical Specifications requires that the channel sensor,
including the primary radiation detector, be exposed to a
radioactive source.
The licensee's amendment request would not change the
definition for source check; however, if the amendment were
approved, the licensee would be free to relax the
definition for source check under its ODCM, provided they
met the criteria that "the over-all level of radiological
effluent control is not reduced." A violation of this
criteria would be a violation of the licensee's Technical
Specification.
The NRR staff have adopted the position that any proposal
by a licensee to relax the definition of source check,
whether through an amendment request or under its ODCM
pursuant to Generic Letter 89-01, is not acceptable without
the licensee providing compensatory measures for the
proposed relaxation. This is necessary because such
changes on measurements can reduce the overall effluent
control. Therefore, the following conditions must be met:
1. If the detector of concern is used as the primary
means of quantifying radionuclides in effluent streams, the
licensee must proved justification on why an alternative
and technically more accurate measurement (e.g. taking grab
samples) is not available. If an alternative measurement
is not available, then detector specific and other
effluent-related information should be provided either in
the ODCM or other means for the staff to evaluate whether
the overall effluent control will be reduced.
2. If the scintillator plastic / photomultiplier type
detector is used only for detecting radiation that
activates the alarm / trip setpoint, relaxation of the
current source check definition should be accompanied by a
commitment from the licensee to provide compensatory
measures to ensure the overall effluent control not be
reduced over time and usage. A commitment by the licensee
to Cross-check and Document the detector scaler count-rate
with the grab sample result (C&D measurement), where
practical, in lieu of the monthly source check measurement,
would be acceptable. In those situations where the C&D
measurement or other comparable measurements are not
practical, the use of the LED light source and/or secondary
check source measurements would be acceptable.
Regulatory references: 10 CFR 50, Regulatory Guide 1.21,
Regulatory Guide 4.15, Technical Specifications
Subject codes: 6.6, 7.3, 12.12
Applicability: Source

