Part 21 Report - 1998-490
LICENSEE: ENTEK IRD INTERNATIONAL CORP. SITE: ENTEK IRD INTERNATIONAL CORP. EN NUMBER:34371 DOCKET: EVENT DATE: 06-09-98 RX TYPE: EVENT TIME: 00:00 VENDORS: NOTIFY DATE: 06-09-98 EMERGENCY CLASS: N/A REGION: 3 STATE: OH TIME: 13:48 OPS OFFICER: CHAUNCEY GOULD 10 CFR SECTION: CCCC 21.21 UNSPECIFIED PARAGRAPH REPORT IN ACCORDANCE WITH 10 CFR PART 21 ENTEK IRD INTERNATIONAL, CORP., IS THE SUPPLIER OF CALIBRATION SERVICES FOR NIST TRACEABILITY OF VIBRATION MEASUREMENTS FOR PORTABLE VIBRATION MEASUREMENT EQUIPMENT. PRIOR TO A MERGER BETWEEN ENTEK SCIENTIFIC AND IRD MECHANALYSIS, INC., IN 1996, AUDITS WERE PERFORMED BY PACIFIC GAS & ELECTRIC AGAINST 10 CFR PART 50, APPENDIX B, AND THERE WAS A QUESTION OF THE ACCURACY OF THE CALIBRATION CERTIFICATES BEING PROVIDED BY IRD MECHANALYSIS, INC. IRD MECHANALYSIS, INC., HAD BEEN PROVIDING CALIBRATION CERTIFICATES INDICATING ACCURACY FOR TRANSDUCERS CALIBRATED ON ITS UNHOLTZ-DICKIE SHAKER TABLES WITH A 4:1 ACCURACY RATIO WHEN IT ACTUALLY HAD A 3.125:1 ACCURACY RATIO. THE COMPANY IS ADVISING LICENSEES TO EVALUATE THE IMPACT OF A CHANGE FROM A 4:1 ACCURACY RATIO TO A 3.125:1 ACCURACY RATIO. (REFER TO THE HOO LOG FOR A CONTACT TELEPHONE NUMBER.)