Event Notification Report for May 28, 2017

U.S. Nuclear Regulatory Commission
Operations Center

EVENT REPORTS FOR
05/27/2017 - 05/28/2017

EVENT NUMBERS
52774

Agreement State
Event Number: 52774
Rep Org: MISSISSIPPI DIV OF RAD HEALTH
Licensee: MS POWER COMPANY
Region: 4
City: DE KALB   State: MS
County:
License #: MS-1062-01
Agreement: Y
Docket:
NRC Notified By: JAYSON MOAK
HQ OPS Officer: JEFF HERRERA
Notification Date: 05/30/2017
Notification Time: 10:19 [ET]
Event Date: 05/28/2017
Event Time: 00:00 [CDT]
Last Update Date: 07/05/2017
Emergency Class: NON EMERGENCY
10 CFR Section:
Person (Organization):
NEIL OKEEFE (R4DO)
NMSS_EVENTS_NOTIFICA (EMAI)
Event Text
AGREEMENT STATE REPORT - STUCK SHUTTER ON A FIXED GAUGE

The following report was received from the Mississippi Department of Health via email:

"Licensee's RSO [Radiation Safety Officer] reported a stuck shutter on a Ronan Model SA8-C10 fixed gauge, Serial No. 1093CP, Source Model CDC.700. This was noticed during preventive maintenance, while performing shutter checks. Licensee notified the manufacturer for service.

"Isotope: Cs-137
Activity: 200 mCi"

* * * UPDATE AT 1524 EDT ON 7/5/17 FROM JAYSON MOAK TO MARK ABRAMOVITZ * * *

The following update was received via e-mail:

"Licensee reported that the gauge had been repaired by the manufacturer and returned to service. The gauge manufacturer's field engineer reported that the gauge shutter was able to be closed with channel lock pliers before the gauge was removed from the vessel. The field engineer reported that the lead ceiling portion of the shutter cavity was tight against the shutter assembly causing the shutter assembly to bind and not close. The field engineer suspects that water leached in behind the lead and froze, thereby pushing the lead deeper into the shutter cavity. The field engineer removed the expanded lead from hitting the shutter. Once the gauge had its seals replaced, the source holder was reassembled. The gauge was then wiped for removable contamination, surveyed at 30 cm and 12 in, and remounted by the field engineer for continued use. The wipe test was negative (less than 0.001uCi) for removable contamination and all fifteen (15) surveys were less than 1 mR/Hr with the gauge shutter open."

Mississippi Report: MS-17001

Notified the R4DO (Farnholtz) and NMSS Resources (via e-mail).