General Information or Other
Event Number: 43970
Rep Org: CALIFORNIA RADIATION CONTROL PRGM
Licensee: UNIVERSITY OF CALIFORNIA BERKELEY
Region: 4
City: BERKELEY   State: CA
County:
License #: 1333-01
Agreement: Y
Docket:
NRC Notified By: MARK GOTTLIEB
HQ OPS Officer: MARK ABRAMOVITZ
Notification Date: 02/07/2008
Notification Time: 19:47 [ET]
Event Date: 08/09/2007
Event Time: 00:00 [PST]
Last Update Date: 02/07/2008
Emergency Class: NON EMERGENCY
10 CFR Section:
Person (Organization):
MICHAEL SHANNON (R4)
Event Text
AGREEMENT STATE - LEAKING GENERAL LICENSED SOURCE FROM A GAS CHROMATOGRAPH

The State provided the following information via email:

"On August 9, 2007, a dry smear of an ECD at UC Berkeley indicated a gross count rate of 5171 cpm. It was determined that this count rate indicated an activity of 0.03 uCi of Nickel-63. The ECD was returned to Varian on August 21, 2007, who determined that the leakage was less than the reportable limit of 0.005 uCi. Varian returned the ECD with the original Nickel-63 foil to UC Berkeley on August 28, 2007.

"On October 16, 2007, the same 15 mCi Nickel-63 foil in the same Electron Capture Detector (ECD) in the Varian gas chromatograph was again found to be leaking. The activity of the leakage was determined to be 0.06 uCi. On October 17, 2007, the use of the gas chromatograph was immediately suspended, and a follow-up survey found that the contamination was limited to the top of the ECD, the exhaust funnel and a section of the exhaust tubing. On October 19, 2007, Varian representatives removed the ECD and transported it to their facility for analysis and repair. No adverse health effects resulted from this incident. UC Berkeley notified the RHB [Radiologic Health Branch] of this matter on November 9, 2007."