Event Notification Report for July 20, 2007

U.S. Nuclear Regulatory Commission
Operations Center

EVENT REPORTS FOR
07/19/2007 - 07/20/2007

EVENT NUMBERS
435104351344303

Other Nuclear Material
Event Number: 43510
Rep Org: MATTINGLY TESTING SERVICES
Licensee: MATTINGLY TESTING SERVICES
Region: 4
City: LAUREL   State: MT
County:
License #: 25-21479-01
Agreement: N
Docket:
NRC Notified By: DENNIS FICEK
HQ OPS Officer: BILL HUFFMAN
Notification Date: 07/20/2007
Notification Time: 21:00 [ET]
Event Date: 07/20/2007
Event Time: 17:00 [MDT]
Last Update Date: 07/20/2007
Emergency Class: NON EMERGENCY
10 CFR Section:
Person (Organization):
ANTHONY GODY (R4)
E. WILLIAM BRACH (FSME)
LEN WERT (R4)
CHUCK CAIN (R4)
VIVIAN CAMPBELL (R4)
Event Text
RADIOGRAPHY CAMERA SOURCE STUCK IN GUIDE TUBE

A manager from the licensee's company called to inform the NRC of a problem involving a radiography camera with a source stuck in the guide tube. The radiography camera, an INC IR-100 with a 97 curie Iridium-192 source, was being used at the Cenex Refinery in Laurel, MT. The camera was perched at an elevated location on an I-beam when the collimator fell and put a sharp bend in the guide tube. The radiographer attempted to retract the source but was unable to get it past the bend. The radiographer left the camera and unretracted source in its work location to minimize his exposure to the source after ensuring that it was well secured.

The radiographer established a 2 millirem boundary around the camera location and then attempted to call the RSO. The radiographer also called a company manager who responded to the site and subsequently contacted the NRC. Currently there is no health and safety concern because the area is cordoned off and the location of the camera is in a part of the refinery that is being expanded (new construction) so there is no concern about refinery worker exposure. The RSO is not at his home and cannot be reached by cell because of the spotty cell coverage in many parts of Montana. The licensee plans to maintain 24 hour around the clock coverage of the boundary and camera until expertise can be brought to the site to recover the source or place it in a condition where it can be removed from its current location without excessive exposure. There are currently 5 employees from the company onsite to maintain this coverage. The licensee will continue to attempt to contact the RSO but the worst case condition may require coverage to be extended until normal working hours on Monday morning (7/23) when the RSO comes back to town. The licensee is also trying to contact the RSO of a competitive radiography company in the area that could lend assistance.

The licensee's representative plans to brief the NRC in approximately 12 hours on the status of their efforts to secure the source.

* * * UPDATE FROM LICENSEE (FICEK) TO HUFFMAN AT 2300 EDT ON 7/20/07 * * *

The licensee was able to fabricate a tool using a 15 foot section of conduit with a vice grip taped to the end. Using this tool, they were able to push the collimator up level with the camera and straighten the bend in the guide tube. The radiographer was then able to crank the source back into the camera. The source is now in its safe shielded position and has been secured in the company truck and the licensee is preparing to leave the site. The licensee did not know what level of exposure the radiographer received during this event but did not believe it was excessive.

R4DO (Gody), R4 (Wert), and NMSS EO (Brach) notified.


General Information or Other
Event Number: 43513
Rep Org: ILLINOIS EMERGENCY MGMT. AGENCY
Licensee: MCNDT LEASING INC
Region: 3
City: DARIEN   State: IL
County:
License #: IL-01875-01
Agreement: Y
Docket:
NRC Notified By: DAREN PERRERO
HQ OPS Officer: JASON KOZAL
Notification Date: 07/23/2007
Notification Time: 15:36 [ET]
Event Date: 07/20/2007
Event Time: 08:00 [CDT]
Last Update Date: 07/23/2007
Emergency Class: NON EMERGENCY
10 CFR Section:
Person (Organization):
RICHARD SKOKOWSKI (R3)
MICHELE BURGESS (FSME)
Event Text
AGREEMENT STATE - STUCK RADIOGRAPHY CAMERA SOURCE

"On the morning of July 20, [DELETED], RSO for McNDT, called. He was taking some radiographs while inside a water tower in Darien, IL and the magnetic cable stand used to hold the film cartridge fell over onto the cable and crushed the guide tube. He reported that he could not retract the source to the shielded position in the camera. The source is a 34 Ci Ir-192 source; the camera is a QSA 660B. The scene was secured and under constant oversight by the RSO. [DELETED] reported that there have been no overexposures or any injuries. [DELETED] estimated that some 6 inches of guide tube used to position the radiography source had been crushed/deformed by equipment that had been used to position the radiography source 8 feet high up against the wall of the tank. The guide tube was then stretched straight along the floor of the water tower which is approximately 50 feet in diameter. The damaged section of the tube was approximately 30 inches from the end of the 14 foot length of guide tube. A 'hot zone' barrier required by regulations was established around the water tower before the radiography job was begun. Dose rates around that zone were less than 60 milliR/hr. The water tower is near the intersection of Plainfield and Cass Avenues in Darien behind a retail shopping area. As the area is a construction area, there is fencing surrounding the work site so control of the scene has been relatively easy to maintain.

"Upon arrival at the scene, the investigators added bags of lead shot to the area where the source was determined to be. The drive cables and crank were then physically secured with the camera and a strong force was applied to the cranking mechanism. This resulted in the successful return of the cable with the Ir-192 source back into the radiography camera. Afterwards, the guide tube was surveyed and found to be free of contamination. The camera ports were likewise monitored with negative removable contamination present. The device (AEA QSA, Inc. model 660B w/ S/N B2020) has been secured and tagged as out of service until it can be returned to the manufacturer for source exchange and recertification. Surface dose rates on the radiography camera were at the expected 5 to 10 milliR/hr. Total exposures to personnel as measured by pocket dosimeters were very low (70 milliR to the RSO, 60 milliR to [DELETED] and 55 milliR to [DELETED]). [DELETED] was advised of our expectations for return of the potentially damaged source cable and the need to send in his own dosimetry in for immediate analysis to confirm the measured exposure prior to our staffs' arrival."

Illinois Item Number: IL070040


General Information or Other
Event Number: 44303
Rep Org: NE DIV OF RADIOACTIVE MATERIALS
Licensee: 'R' PLACE
Region: 4
City: HORDVILLE   State: NE
County:
License #: GL0683
Agreement: Y
Docket:
NRC Notified By: TRUDY HILL
HQ OPS Officer: JASON KOZAL
Notification Date: 06/18/2008
Notification Time: 08:28 [ET]
Event Date: 07/20/2007
Event Time: 00:00 [CDT]
Last Update Date: 06/18/2008
Emergency Class: NON EMERGENCY
10 CFR Section:
Person (Organization):
THOMAS FARNHOLTZ (R4)
MICHELE BURGESS (FSME)
ILTAB VIA E-MAIL
Event Text
AGREEMENT STATE REPORT - LOST TRITIUM EXIT SIGNS

The state provided the following information via e-mail:

"The Department contacted the general licensee because the annual fee for the general license was overdue for two self illuminating signs. The signs were from Best Lighting (Model SLTURW10)(Serial # 166473 and 166474) with 7.03 curies [each] of H3 (in June of 2006). The general licensee hired an individual to remove the signs from the dining room and put them into the storage room in the cafÚ during July of 2007.

"The general licensee went to the storage room in the second week of August 2007 to retrieve some stored items. Items were missing and additionally the two exit signs were also missing. The general licensee tried to contact the man who she hired to move the supplies and equipment. She did not find a phone listing. She sent him a letter to the address he had given her and the letter was returned as no such address."

Nebraska event number: NE080003.

THIS MATERIAL EVENT CONTAINS A "LESS THAN CAT 3" LEVEL OF RADIOACTIVE MATERIAL

Sources that are "Less than IAEA Category 3 sources," are either sources that are very unlikely to cause permanent injury to individuals or contain a very small amount of radioactive material that would not cause any permanent injury. Some of these sources, such as moisture density gauges or thickness gauges that are Category 4, the amount of unshielded radioactive material, if not safely managed or securely protected, could possibly - although it is unlikely - temporarily injure someone who handled it or were otherwise in contact with it, or who were close to it for a period of many weeks.

This source is not amongst those sources or devices identified by the IAEA Code of Conduct for the Safety & Security of Radioactive Sources to be of concern from a radiological standpoint. Therefore is it being categorized as a less than Category 3 source