Event Notification Report for July 15, 2007
U.S. Nuclear Regulatory Commission
Operations Center
EVENT REPORTS FOR
07/14/2007 - 07/15/2007
General Information or Other
Event Number: 43493
Rep Org: CALIFORNIA RADIATION CONTROL PRGM
Licensee: CITY OF HOPE
Region: 4
City: DUARTE State: CA
County:
License #: 0307-19
Agreement: Y
Docket:
NRC Notified By: K. KAUFMAN
HQ OPS Officer: MARK ABRAMOVITZ
Licensee: CITY OF HOPE
Region: 4
City: DUARTE State: CA
County:
License #: 0307-19
Agreement: Y
Docket:
NRC Notified By: K. KAUFMAN
HQ OPS Officer: MARK ABRAMOVITZ
Notification Date: 07/16/2007
Notification Time: 20:38 [ET]
Event Date: 07/15/2007
Event Time: 22:00 [PDT]
Last Update Date: 07/16/2007
Notification Time: 20:38 [ET]
Event Date: 07/15/2007
Event Time: 22:00 [PDT]
Last Update Date: 07/16/2007
Emergency Class: NON EMERGENCY
10 CFR Section:
10 CFR Section:
Person (Organization):
GREG PICK (R4)
JOSEPH HOLONICH (FSME)
GREG PICK (R4)
JOSEPH HOLONICH (FSME)
AGREEMENT STATE REPORT - UNPLANNED EXPOSURE DURING CALIBRATION TEST
The State provided the following information via facsimile:
A doctor was performing a monthly calibration test and position verification test. During the test, an error message was received using a dummy source. The doctor entered the room with a survey meter which showed no elevated radiation reading. The emergency handwheel was used to retract the dummy wire. After about 12 turns, the room radiation alarms initiated and the doctor left the room. Troubleshooting disclosed that the source had become exposed and the doctor received approximately 75 Mr exposure.
Source: Ir-192, 5.2 Curies
Varian Medical Systems model Varisource HDR
The State provided the following information via facsimile:
A doctor was performing a monthly calibration test and position verification test. During the test, an error message was received using a dummy source. The doctor entered the room with a survey meter which showed no elevated radiation reading. The emergency handwheel was used to retract the dummy wire. After about 12 turns, the room radiation alarms initiated and the doctor left the room. Troubleshooting disclosed that the source had become exposed and the doctor received approximately 75 Mr exposure.
Source: Ir-192, 5.2 Curies
Varian Medical Systems model Varisource HDR
General Information or Other
Event Number: 43647
Rep Org: VARIAN MEDICAL SYSTEMS
Licensee: VARIAN MEDICAL SYSTEMS
Region: 1
City: CHARLOTTESVILLE State: VA
County:
License #: 45-30957-01
Agreement: N
Docket:
NRC Notified By: RICHARD PICCOLO
HQ OPS Officer: JOE O'HARA
Licensee: VARIAN MEDICAL SYSTEMS
Region: 1
City: CHARLOTTESVILLE State: VA
County:
License #: 45-30957-01
Agreement: N
Docket:
NRC Notified By: RICHARD PICCOLO
HQ OPS Officer: JOE O'HARA
Notification Date: 09/17/2007
Notification Time: 17:19 [ET]
Event Date: 07/15/2007
Event Time: 00:00 [EDT]
Last Update Date: 09/17/2007
Notification Time: 17:19 [ET]
Event Date: 07/15/2007
Event Time: 00:00 [EDT]
Last Update Date: 09/17/2007
Emergency Class: NON EMERGENCY
10 CFR Section:
21.21 - UNSPECIFIED PARAGRAPH
10 CFR Section:
21.21 - UNSPECIFIED PARAGRAPH
Person (Organization):
PAUL KROHN (R1)
JOEL MUNDAY (R2)
HIRONORI PETERSON (R3)
VIVIAN CAMPBELL (R4)
PATRICE BUBAR (FSME)
PAUL KROHN (R1)
JOEL MUNDAY (R2)
HIRONORI PETERSON (R3)
VIVIAN CAMPBELL (R4)
PATRICE BUBAR (FSME)
ACTIVE SOURCE CAN BECOME DISLODGED IN HDR AFTERLOADER DEVICE
The following information was received via fax:
Manufacturer submitted this notification based on an event involving a High Dose Rate (HDR) Afterloader that occurred at City of Hope Hospital in Duarte, California on July 15, 2007 [See EN# 43493]
"Identification of the facility, the activity, or the basic component supplied for such facility or such activity within the United States which fails to comply or contains a defect- VariSource HDR Afterloader model 200 and VariSource HDR Afterloader model ID
"Identification of the firm constructing the facility or supplying the basic component which fails to comply or contains a defect:
The VariSource HDR Afterloader is manufactured by: Varian-TEM Ltd., Gatwick Road, Crawley, West Sussex RH102RG, United Kingdom
"Nature of the defector failure to comply and the safety hazard which is created or could be created by such defect or failure to comply:
The active source can become dislodged from the internal tungsten safe if the emergency source retract hand crank is used when the active source is safely parked in the internal shield. (1) This event can only occur if the operator turns the emergency hand crank while the source is safely parked. (2) The emergency source retract hand crank is to be used only when the active source is in the out or exposed position. It is not designed, nor intended to be operated when the source is safely parked. (3) The emergency hand crank is only connected to the active source wire. It is not connected to the dummy source and its operation has no effect on the dummy source."
Customer Technical Bulletin, CTB-VS-366A, "Clarification on the use of the emergency retract hand wheel to prevent accidental exposure" was issued to all users of the VariSource HDR unit on April 27, 2004. This CTB will be reissued to all domestic customers. A warning label addressing the proper use of the emergency hand crank was attached to all VariSource HDR units in 2004. A new warning label is being developed that will restate the proper use of the hand crank. Also a formal revision was made to user training to emphasize that this event is possible if the user makes an error by operating the emergency hand crank when the active source wire is safely parked in the tungsten shield.
There are 186 VariSource HDR units in the United States and a list was provided to the NRC.
The following information was received via fax:
Manufacturer submitted this notification based on an event involving a High Dose Rate (HDR) Afterloader that occurred at City of Hope Hospital in Duarte, California on July 15, 2007 [See EN# 43493]
"Identification of the facility, the activity, or the basic component supplied for such facility or such activity within the United States which fails to comply or contains a defect- VariSource HDR Afterloader model 200 and VariSource HDR Afterloader model ID
"Identification of the firm constructing the facility or supplying the basic component which fails to comply or contains a defect:
The VariSource HDR Afterloader is manufactured by: Varian-TEM Ltd., Gatwick Road, Crawley, West Sussex RH102RG, United Kingdom
"Nature of the defector failure to comply and the safety hazard which is created or could be created by such defect or failure to comply:
The active source can become dislodged from the internal tungsten safe if the emergency source retract hand crank is used when the active source is safely parked in the internal shield. (1) This event can only occur if the operator turns the emergency hand crank while the source is safely parked. (2) The emergency source retract hand crank is to be used only when the active source is in the out or exposed position. It is not designed, nor intended to be operated when the source is safely parked. (3) The emergency hand crank is only connected to the active source wire. It is not connected to the dummy source and its operation has no effect on the dummy source."
Customer Technical Bulletin, CTB-VS-366A, "Clarification on the use of the emergency retract hand wheel to prevent accidental exposure" was issued to all users of the VariSource HDR unit on April 27, 2004. This CTB will be reissued to all domestic customers. A warning label addressing the proper use of the emergency hand crank was attached to all VariSource HDR units in 2004. A new warning label is being developed that will restate the proper use of the hand crank. Also a formal revision was made to user training to emphasize that this event is possible if the user makes an error by operating the emergency hand crank when the active source wire is safely parked in the tungsten shield.
There are 186 VariSource HDR units in the United States and a list was provided to the NRC.