Skip to main content

Event Notification Report for January 09, 2007

U.S. Nuclear Regulatory Commission
Operations Center

EVENT REPORTS FOR
01/08/2007 - 01/09/2007

EVENT NUMBERS
4308743084

General Information or Other
Event Number: 43087
Rep Org: SC DIV OF HEALTH & ENV CONTROL
Licensee: CARE ALLIANCE HEALTH SERVICES ROPER HOSPITAL
Region: 1
City: CHARLESTON   State: SC
County:
License #: 646
Agreement: Y
Docket:
NRC Notified By: JAMES PETERSON
HQ OPS Officer: MARK ABRAMOVITZ
Notification Date: 01/10/2007
Notification Time: 17:30 [ET]
Event Date: 01/09/2007
Event Time: 00:00 [EST]
Last Update Date: 01/10/2007
Emergency Class: NON EMERGENCY
10 CFR Section:
Person (Organization):
RONALD BELLAMY (R1)
JOSEPH GIITTER (NMSS)
Event Text
AGREEMENT STATE REPORT - POSSIBLE MEDICAL EVENT - BRACHYTHERAPY

The State provided the following information via facsimile:

"Licensee has notified State Agency within 24 hours.

"A Mick applicator malfunctioned during the planned treatment of a prostate patient. The patient was scheduled to receive I-125 seed implants totaling 11 millicuries at 0.25 millicuries per seed. A total of 8.25 millicuries were implanted when the malfunction occurred. This will most likely result in the total dose delivered differing from prescribed dose by 20 percent or more. The dose information was not currently available but will follow in NMED reporting. The seeds not implanted have all been accounted for and have been placed in storage. The licensee will provide a written report within 15 days. No further information is available at this time."

SC Event Report: SC070001

A "Medical Event" may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.


General Information or Other
Event Number: 43084
Rep Org: TEXAS DEPARTMENT OF HEALTH
Licensee: H&G INSPECTION COMPANY
Region: 4
City: HOUSTON   State: TX
County:
License #: L02181
Agreement: Y
Docket:
NRC Notified By: LATISCHA HANSON
HQ OPS Officer: JEFF ROTTON
Notification Date: 01/09/2007
Notification Time: 18:17 [ET]
Event Date: 01/09/2007
Event Time: 00:00 [CST]
Last Update Date: 01/09/2007
Emergency Class: NON EMERGENCY
10 CFR Section:
Person (Organization):
BLAIR SPITZBERG (R4)
JOSEPH GIITTER (NMSS)
Event Text
AGREEMENT STATE REPORT - POTENTIAL OVEREXPOSURE DISCOVERED AFTER BADGE PROCESSING

"The agency [Texas Department of State Health Services (DSHS)] returned a telephone call at 5 p.m. [on 01/09/07] to the licensee, in regards to a voicemail the licensee's RSO left on another program's telephone. The personnel answering the telephone at the facility stated the RSO had just left for the day & gave [the state] partial information that the facility had received the personnel monitoring report for 11/10/06-12/06/06 that showed a radiographer's dose to be 697,408 (mR) -Deep dose. The employee stated that the monitoring processing company was going to repeat the tests again to confirm the dose. The licensee was waiting to hear from DSHS/Radiation Control (RC) on 1/09/07, to decide what to do. They were not sure if they should do blood work, etc.

"DSHS/RC advised the employee to 1) Have the RSO call DSHS/RC Incident Investigation (IIP) personnel first thing in the morning on 1/10/07; 2) Have the RSO perform an inquiry with the radiographer to assess the events during the monitoring period in question; 3) Have the RSO look at personnel monitoring records for co-workers on the same job(s) as the radiographer under review to compare their dosage during this time period; 4) Have RSO check utilization logs during this period to assess what specific equipment was used & the job details; 5) DSHS/RC requested a fax copy of the personnel monitoring report [Deleted]. The employee agreed to comply. DSHS/RC will follow-up first thing 1/10/07 with the licensee's RSO to obtain complete information. Additional clarification/corrective information may be submitted by DSHS/RC after interview with licensee's RSO. "

Texas Incident number: I-8383