Event Notification Report for July 10, 2006
U.S. Nuclear Regulatory Commission
Operations Center
EVENT REPORTS FOR
07/09/2006 - 07/10/2006
General Information or Other
Event Number: 42729
Rep Org: OHIO BUREAU OF RADIATION PROTECTION
Licensee: AKRON GENERAL MEDICAL CENTER
Region: 3
City: AKRON State: OH
County:
License #: 02120780000
Agreement: Y
Docket:
NRC Notified By: MARK LIGHT
HQ OPS Officer: MARK ABRAMOVITZ
Licensee: AKRON GENERAL MEDICAL CENTER
Region: 3
City: AKRON State: OH
County:
License #: 02120780000
Agreement: Y
Docket:
NRC Notified By: MARK LIGHT
HQ OPS Officer: MARK ABRAMOVITZ
Notification Date: 07/26/2006
Notification Time: 10:48 [ET]
Event Date: 07/10/2006
Event Time: 00:00 [EDT]
Last Update Date: 07/26/2006
Notification Time: 10:48 [ET]
Event Date: 07/10/2006
Event Time: 00:00 [EDT]
Last Update Date: 07/26/2006
Emergency Class: NON EMERGENCY
10 CFR Section:
10 CFR Section:
Person (Organization):
ROGER LANKSBURY (R3)
GREG MORELL (NMSS)
ROGER LANKSBURY (R3)
GREG MORELL (NMSS)
AGREEMENT STATE REPORT - MEDICAL OVERDOSE
The following information was provided by the state related to prostate brachytherapy which was performed on 7/10/2006 using I-125 seeds.
"Prostate seed implant plans are specified in kerma units (U) by default in our computer planning system. However, the ordering of seeds is specified in mCi [milliCuries]. In this instance the default seed strength was not changed to mCi and a plan was developed for [the patient] to receive 111 seeds of an activity of 0.394 U per seed when 0.394 mCi was desired. The order form was completed for 111 seeds of the expected (not planned) activity of 0.394 mCi per seed. The result was an implant with seeds of activity 27% higher than planned."
The overdose was noticed by the hospital on 7/12/2006. The physician, the patient, and the State of Ohio were notified on 7/13/2006. The State inspected the facility on 7/18/2006.
The following information was provided by the state related to prostate brachytherapy which was performed on 7/10/2006 using I-125 seeds.
"Prostate seed implant plans are specified in kerma units (U) by default in our computer planning system. However, the ordering of seeds is specified in mCi [milliCuries]. In this instance the default seed strength was not changed to mCi and a plan was developed for [the patient] to receive 111 seeds of an activity of 0.394 U per seed when 0.394 mCi was desired. The order form was completed for 111 seeds of the expected (not planned) activity of 0.394 mCi per seed. The result was an implant with seeds of activity 27% higher than planned."
The overdose was noticed by the hospital on 7/12/2006. The physician, the patient, and the State of Ohio were notified on 7/13/2006. The State inspected the facility on 7/18/2006.
Other Nuclear Material
Event Number: 42695
Rep Org: READING SITE CONTRACTORS
Licensee: WINDSOR SERVICES
Region: 1
City: READING State: PA
County:
License #: 37-18494-01
Agreement: N
Docket:
NRC Notified By: STAN LEGANZA
HQ OPS Officer: JOE O'HARA
Licensee: WINDSOR SERVICES
Region: 1
City: READING State: PA
County:
License #: 37-18494-01
Agreement: N
Docket:
NRC Notified By: STAN LEGANZA
HQ OPS Officer: JOE O'HARA
Notification Date: 07/10/2006
Notification Time: 10:23 [ET]
Event Date: 07/10/2006
Event Time: 09:41 [EDT]
Last Update Date: 07/10/2006
Notification Time: 10:23 [ET]
Event Date: 07/10/2006
Event Time: 09:41 [EDT]
Last Update Date: 07/10/2006
Emergency Class: NON EMERGENCY
10 CFR Section:
10 CFR Section:
Person (Organization):
GLENN MEYER (R1)
GREG MORELL (NMSS)
GLENN MEYER (R1)
GREG MORELL (NMSS)
DAMAGED MOISTURE DENSITY GAUGE
A Troxler Model 4640B moisture density gauge was damaged when it was run over by an asphalt roller at a construction site located approximately 150 feet south of the intersection of Route 10 (Morgantown Road) and Upland Avenue in Reading, PA. A perimeter has been established around the gauge and the licensee's RSO has been notified and is enroute to the site. The gauge contains a Cs-137 source with a 8 millicurie activity level. Subsequent discussions with a Berks County Emergency Management official indicates that radiation levels are normal and that the source is intact inside the gauge. Berks County Emergency Management is awaiting authorization from PEMA to remove the gauge from the earth and perform surveys underneath the gauge.
A Troxler Model 4640B moisture density gauge was damaged when it was run over by an asphalt roller at a construction site located approximately 150 feet south of the intersection of Route 10 (Morgantown Road) and Upland Avenue in Reading, PA. A perimeter has been established around the gauge and the licensee's RSO has been notified and is enroute to the site. The gauge contains a Cs-137 source with a 8 millicurie activity level. Subsequent discussions with a Berks County Emergency Management official indicates that radiation levels are normal and that the source is intact inside the gauge. Berks County Emergency Management is awaiting authorization from PEMA to remove the gauge from the earth and perform surveys underneath the gauge.