Event Notification Report for February 12, 2004
U.S. Nuclear Regulatory Commission
Operations Center
EVENT REPORTS FOR
02/11/2004 - 02/12/2004
Power Reactor
Event Number: 40516
Facility: CLINTON
Region: 3 State: IL
Unit: [1] [] []
RX Type: [1] GE-6
NRC Notified By: KEN LEFFEL
HQ OPS Officer: ERIC THOMAS
Region: 3 State: IL
Unit: [1] [] []
RX Type: [1] GE-6
NRC Notified By: KEN LEFFEL
HQ OPS Officer: ERIC THOMAS
Notification Date: 02/12/2004
Notification Time: 21:14 [ET]
Event Date: 02/12/2004
Event Time: 17:45 [CST]
Last Update Date: 02/12/2004
Notification Time: 21:14 [ET]
Event Date: 02/12/2004
Event Time: 17:45 [CST]
Last Update Date: 02/12/2004
Emergency Class: NON EMERGENCY
10 CFR Section:
50.72(b)(3)(xiii) - LOSS COMM/ASMT/RESPONSE
10 CFR Section:
50.72(b)(3)(xiii) - LOSS COMM/ASMT/RESPONSE
Person (Organization):
ERIC DUNCAN (R3)
ERIC DUNCAN (R3)
| Unit | SCRAM Code | RX Crit | Initial PWR | Initial RX Mode | Current PWR | Current RX Mode |
|---|---|---|---|---|---|---|
| 1 | N | N | 0 | Refueling | 0 | Refueling |
SAFETY PARAMETER DISPLAY SYSTEM (SPDS) OUT OF SERVICE FOR MAINTENANCE.
"The Safety Parameter Display System (SPDS) was removed from service on February 12, 2004, at 1745 CST, to support planned maintenance on uninterruptible power supply (UPS) 1B. SPDS will be out of service for greater than eight hours. The UPS 1B provides power to the Nuclear Steam Supply (NSS) Computer, which provides the computer points for SPDS. The UPS 1B outage is scheduled for approximately twenty-eight (28) hours. Operators are taking compensatory measures for monitoring parameters.
"10CFR50.72(b)(3)(xiii) states 'the licensee shall notify the NRC as soon as practical and in all cases within 8 hours of the occurrence of any event that results in the major loss of emergency assessment capability.'
"The loss of SPDS is considered a major loss of safety assessment capability. The site defines a 'major loss' of SPDS, 'when the SPDS function is not available in the control room for greater than eight hours.' This is consistent with NUREG-1022, EVENT REPORTING GUIDELINES 10CFR 50.72 and 50.73 section 3.2.13, Loss of Emergency Preparedness Capabilities.
"SPDS is expected to be restored to service on February 13, 2004. The plant is in a refueling outage. All other plant conditions are normal in support of the refueling outage. The NRC Resident Inspector has been notified."
"The Safety Parameter Display System (SPDS) was removed from service on February 12, 2004, at 1745 CST, to support planned maintenance on uninterruptible power supply (UPS) 1B. SPDS will be out of service for greater than eight hours. The UPS 1B provides power to the Nuclear Steam Supply (NSS) Computer, which provides the computer points for SPDS. The UPS 1B outage is scheduled for approximately twenty-eight (28) hours. Operators are taking compensatory measures for monitoring parameters.
"10CFR50.72(b)(3)(xiii) states 'the licensee shall notify the NRC as soon as practical and in all cases within 8 hours of the occurrence of any event that results in the major loss of emergency assessment capability.'
"The loss of SPDS is considered a major loss of safety assessment capability. The site defines a 'major loss' of SPDS, 'when the SPDS function is not available in the control room for greater than eight hours.' This is consistent with NUREG-1022, EVENT REPORTING GUIDELINES 10CFR 50.72 and 50.73 section 3.2.13, Loss of Emergency Preparedness Capabilities.
"SPDS is expected to be restored to service on February 13, 2004. The plant is in a refueling outage. All other plant conditions are normal in support of the refueling outage. The NRC Resident Inspector has been notified."
General Information or Other
Event Number: 40668
Rep Org: MA RADIATION CONTROL PROGRAM
Licensee: NITON LLC
Region: 1
City: BILLERICA State: MA
County:
License #:
Agreement: Y
Docket:
NRC Notified By: JOHN SUMARES
HQ OPS Officer: STEVE SANDIN
Licensee: NITON LLC
Region: 1
City: BILLERICA State: MA
County:
License #:
Agreement: Y
Docket:
NRC Notified By: JOHN SUMARES
HQ OPS Officer: STEVE SANDIN
Notification Date: 04/12/2004
Notification Time: 09:17 [ET]
Event Date: 02/12/2004
Event Time: 00:00 [EDT]
Last Update Date: 04/12/2004
Notification Time: 09:17 [ET]
Event Date: 02/12/2004
Event Time: 00:00 [EDT]
Last Update Date: 04/12/2004
Emergency Class: NON EMERGENCY
10 CFR Section:
10 CFR Section:
Person (Organization):
LAWRENCE DOERFLEIN (R1)
DALE POWERS (R4)
TOM ESSIG (NMSS)
LAWRENCE DOERFLEIN (R1)
DALE POWERS (R4)
TOM ESSIG (NMSS)
AGREEMENT STATE REPORT INVOLVING SAFETY EQUIPMENT FAILURE
On 2/12/04, a Niton XRF Portable Analyzer (model Xli 800 - S/N 5089) shutter failed to close when the device was placed in the off mode. The incident occurred at the Pearl Harbor Naval Shipyard in Honolulu, HI and did not involve any significant radiation exposures to personnel. The device was packaged with temporary shielding and returned to Niton for evaluation.
Niton determined that the shutter malfunction was the result of a loose set screw which connects the shutter to the motor. A component change to the Xli 800 series models was made that involves upgrading to a nylock set screw. This was the first instance of this type of failure. Instruments returned to Niton for resourcing or software upgrade will be modified in this manner to prevent shutter failure.
This particular device contains a 30 millicurie Americium-241 source and is used for rapid identification and chemistry of metal alloy composition.
On 2/12/04, a Niton XRF Portable Analyzer (model Xli 800 - S/N 5089) shutter failed to close when the device was placed in the off mode. The incident occurred at the Pearl Harbor Naval Shipyard in Honolulu, HI and did not involve any significant radiation exposures to personnel. The device was packaged with temporary shielding and returned to Niton for evaluation.
Niton determined that the shutter malfunction was the result of a loose set screw which connects the shutter to the motor. A component change to the Xli 800 series models was made that involves upgrading to a nylock set screw. This was the first instance of this type of failure. Instruments returned to Niton for resourcing or software upgrade will be modified in this manner to prevent shutter failure.
This particular device contains a 30 millicurie Americium-241 source and is used for rapid identification and chemistry of metal alloy composition.