Event Notification Report for March 08, 2001

U.S. Nuclear Regulatory Commission
Operations Center

EVENT REPORTS FOR
03/07/2001 - 03/08/2001

EVENT NUMBERS
378143782037955

Power Reactor
Event Number: 37814
Facility: PILGRIM
Region: 1     State: MA
Unit: [1] [] []
RX Type: [1] GE-3
NRC Notified By: JOHN NEAL
HQ OPS Officer: STEVE SANDIN
Notification Date: 03/08/2001
Notification Time: 11:44 [ET]
Event Date: 03/08/2001
Event Time: 09:59 [EST]
Last Update Date: 03/08/2001
Emergency Class: NON EMERGENCY
10 CFR Section:
26.73 - FITNESS FOR DUTY
Person (Organization):
RONALD BELLAMY (R1)
ROBERT DENNIG (NRR)
NADER MAMISH (IRO)
Power Reactor Unit Info
Unit SCRAM Code RX Crit Initial PWR Initial RX Mode Current PWR Current RX Mode
1 N Y 100 Power Operation 100 Power Operation
Event Text
FITNESS FOR DUTY REPORT INVOLVING A SUPERVISOR

A non-licensed supervisor was determined to be under the influence of alcohol during a random test. The employee's access has been terminated. The licensee will inform the NRC resident inspector. Contact the Operations Center for additional details.


Hospital
Event Number: 37820
Rep Org: ILLINOIS DEPT OF NUCLEAR SAFETY
Licensee: LOYOLA UNIVERSITY MEDICAL CENTER
Region: 3
City: MAYWOOD   State: IL
County:
License #: IL-01131-02
Agreement: Y
Docket:
NRC Notified By: JOE KLINGER
HQ OPS Officer: DOUG WEAVER
Notification Date: 03/09/2001
Notification Time: 15:57 [ET]
Event Date: 03/08/2001
Event Time: 12:00 [CST]
Last Update Date: 03/09/2001
Emergency Class: NON EMERGENCY
10 CFR Section:
Person (Organization):
JOHN MADERA (R3)
JOHN GREEVES (NMSS)
Event Text
MEDICAL MISADMINISTRATION

The following text was taken from an email sent to the NRC's Operations Center from the State of Illinois.

The RSO at Loyola University Medical Center, IL-01131-02, Maywood, IL, called at approximately 12:30 p.m., 3/9/01, to report a reportable event involving the Cordis Intravascular Brachytherapy (IVB)
system. He stated that two patients undergoing cardiac therapy on the previous day, had been treated with the Ir-192 seeds to prevent restenosis in a stent. The prescribed dose was 8 gray (800 rads) to the tunica media
layer of the heart. However, the calculated dose to one of the patients was12.5 gray (1250 rads) (56% overdose) and 14.3 gray (1,430 rads) (79% overdose).

The cause of the problem appears to be due to a new method for this institution to determine the actual dose to the intended organ. In the past, dwell times were based on 14 gray being delivered at 2 mm distance from the radioactive sources. This new procedure requires an intravascular ultrasound device to be used and measurements from the intended treatment portion of the heart and the sources are determined. These measurements then determine the dwell time for the prescribed dose. This new measurement method and possible communication problems while taking the measurements and using the measurements contributed to the problem. No further use of this system will be conducted until these problems are resolved and the Radiation Safety Committee and the Department are satisfied that there will be no recurrence.

.


Power Reactor
Event Number: 37955
Facility: LIMERICK
Region: 1     State: PA
Unit: [] [2] []
RX Type: [1] GE-4,[2] GE-4
NRC Notified By: PETE ORPHANOS
HQ OPS Officer: STEVE SANDIN
Notification Date: 04/30/2001
Notification Time: 12:54 [ET]
Event Date: 03/08/2001
Event Time: 08:45 [EDT]
Last Update Date: 04/30/2001
Emergency Class: NON EMERGENCY
10 CFR Section:
50.73(a)(1) - INVALID SPECIF SYSTEM ACTUATION
Person (Organization):
HAROLD GRAY (R1)
Power Reactor Unit Info
Unit SCRAM Code RX Crit Initial PWR Initial RX Mode Current PWR Current RX Mode
2 N Y 88 Power Operation 88 Power Operation
Event Text
INVALID PRIMARY CONTAINMENT ISOLATION SYSTEM ACTUATION CAUSED BY SETPOINT DRIFT

"This 60-day optional report, as allowed by 10 CFR 50.73(a)(1), is being made under the reporting requirement in 10CFR50.73(a)(2)(iv)(A) to describe an invalid actuation of a specific system, namely Primary Containment Isolation System (PCIS).

"On March 8, 2001 at 08:45 hours the main control room received a 2A Uninterruptible Power Supply (UPS)/Reactor Protection System (RPS) trouble alarm and verified the following Engineered Safety Features (ESF) isolations: Drywell Chilled Water (DWCW), Reactor Enclosure Cooling Water (RECW), Instrument Gas and Containment Leak Detector.

"The control room also received an electro-hydraulic control (EHC) electrical malfunction alarm with a loss of cooling fans and inaccurate indication of stop valve and combined intercept valves. Additionally, some cabinet fans, 60 Hz solenoid valves and the bypass jack were also lost.

"These expected isolations were bypassed by procedure for the given condition, and were subsequently restored to their normal operating position. The isolations were due to an invalid signal resulting from the failure of a circuit board trim potentiometer for the undervoltage relay (27-AY24801) of the 2A RPS breaker (2A-C248-52). Failure analysis of the undervoltage relay (ABB Model No. 211T4175F) identified that the trim potentiometer was heat sensitive (manufacturing defect) which caused the relay setpoint to drift up to the actual line voltage, tripping the breaker.

"The 2A RPS undervoltage relay was replaced and the isolations were reset at 15:39 hours.

"The impact on the plant from this event was minimal. The selective isolations and equipment losses during the transient did not result in a plant transient, and the equipment performed as expected.

"This event has been entered into the site-specific corrective action program for resolution and generic implications. Industry operational experience (CE) and vendor database searches identified no related issues for this model relay.

"The NRC site resident has been informed."