Home > NRC Library > ADAMS Public Documents
Partial response to FOIA request for all correspondence w/Lixi,Inc re certificate of registration & safety analysis summary for Device NR-422-D-101-S.Forwards documents listed in App A.
Accession Number: ML20106D429
Date Released: Sunday, October 27, 2024
Package Contents
The following links on this page are to Adobe Portable Document Format (PDF) files. To obtain a free viewer for displaying this format, see our Plugins, Viewers, and Other Tools.
- ML20106D424 - Partial response to FOIA request for all correspondence w/Lixi,Inc re certificate of registration & safety analysis summary for Device NR-422-D-101-S.Forwards documents listed in App A. (5 page(s), 7/25/1984)
- ML20106D462 - Forwards drawing for Lixiscope & training course outline, per 801030 telcon. (5 page(s), 10/30/1980)
- ML20106D715 - Part 21 rept re use of unapproved mechanical foot control to turn Lixiscope on/off.Info from distributor re approval for mod unavailable. (2 page(s), 6/1/1984)
- ML20106D628 - Discusses problems found w/operation of on/off switch of Lixiscope.Use of hydraulic sys design requires careful attention to detail.Malfunction of switch poses dangerous situation of large doses to hands. (2 page(s), 4/26/1984)
- ML20106D674 - Responds to 840504 memo re malfunction of Lixiscope,Model LS-80-001 under License 29-21464-01.Unit will be recalled by vendor & examined to determine cause of failure of on/off mechanism. (2 page(s), 5/16/1984)
- ML20106D758 - Responds to inquiry re malfunction of Lixiscope.Unit discussed had neoprene instead of teflon O-rings w/tendency to stick.No units w/neoprene O-ring design distributed into field & no sticking complaints received w/current design. (63 page(s), 7/6/1984)
- ML20106D514 - Responds to 830321 request for addl info re bench mount Lixiscope Model LS-82.Proper use,handling & storage covered in user industrial training. (14 page(s), 3/25/1983)
- ML20106D569 - Forwards GL Kasyk 840426 ltr describing operating problems w/Lixiscope & possible serious design deficiencies in off/on mechanism.Response requested by 840608. (1 page(s), 5/4/1984)