NRC Logo

Home > NRC Library > ADAMS Public Documents

Safety insp on 931203 of license 48-24533-01.Insp findings: licensee failed to test geometry dependence of dose calibrator (CRC-10R) following repair of unit per 10CFR35.50(c).

Accession Number: ML20063G456

Date Released: Friday, October 25, 2024

Package Contents

The following links on this page are to Adobe Portable Document Format (PDF) files. To obtain a free viewer for displaying this format, see our Plugins, Viewers, and Other Tools.

Privacy Policy | Site Disclaimer