Table I.4, Calculational Method, Part 1---Hourly and Annual Dose Received from Gauge 2
Step No. Description GAUGE 2
Input Data
GAUGE 2
Results
1 Dose received in an hour at known distance from gauge (e.g., from manufacturer's data), in mrem/hr 8 8
2 Square of the distance (ft) at which the Step 1 rate was measured, in ft2 (1) 2 1
3 Square of the distance (ft) from the gauge to the secretary's desk in an unrestricted area, in ft2 (12) 2 144
4 Multiply the results of Step 1 by the results of Step 2 (this is an intermediate result) 8 x 1 = 8
5 Divide the result of Step 4 by the result of Step 3 to calculate dose received in an hour by an individual at the secretary's desk, HOURLY DOSE RECEIVED FROM GAUGE 2, in mrem in an hour 8/144 = .056
6 Multiply the result of Step 5 by 24 hr/d x 365 d/yr = MAXIMUM ANNUAL DOSE RECEIVED FROM GAUGE 2, in mrem in a year 0.056 x 24 x 365 =0.056 x 8760 = 491