| NRC INFORMATION NOTICE 2000-16: | POTENTIAL HAZARDS DUE TO VOLATILIZATION OF RADIONUCLIDES |
Addressees:
All U.S. Nuclear Regulatory Commission (NRC) licensees that process unsealed byproduct material.
Purpose:
NRC is issuing this information notice (IN) to alert addressees to the potential hazards associated with the volatilization of radiochemicals and/or radiopharmaceuticals if containment is breached during chemical or physical processing.
The incident described below involves the volatilization of technetium-99m (Tc-99m) during the manufacture of a cardiac imaging agent in a radiopharmacy. However, licensees should be aware of the potential hazards posed by the volatilization of other radionuclides under similar conditions and ensure that their emergency procedures adequately address those hazards.
It is expected that recipients will review this information for applicability to their operations and consider actions, as appropriate. However, suggestions contained in this information notice are not new NRC requirements; therefore, no specific action nor written response is required.
Background:
IN 95-07, issued on January 27, 1995, described the potential for cracking vials and significant contamination when medical and radiopharmacy licensees heat vials of Cardiolite, a cardiac imaging agent, as part of the process to label the pharmaceutical with Tc-99m. The IN described incidents in which vials cracked during the heating phase of the tagging process of Cardiolite, and the subsequent volatilization of the Tc-99m which resulted in significant facility contamination.
Description of Circumstances:
In August of 1999, NRC conducted a special inspection to review the circumstances of a similar accident at a radiopharmacy. The incident occurred while the licensee was using a heating block to process 35 gigabecquerels (950 millicuries) of Tc-99m in 1.3 milliliters of Cardiolite solution. The vial ruptured, and the heat from the block caused the liquid to volatilize, spreading contamination in the laboratory as well as to unrestricted areas throughout the pharmacy.
The licensee's employees did not immediately recognize the volatilization induced spread of the contamination, and continued to work in the laboratory. They did not follow their emergency procedure, which required evacuation of the laboratory in case of a major spill [defined as a spill involving more than 3.7 gigabecquerels (100 millicuries) of Tc-99m]. In addition, the ventilation system was not shut down, causing circulation of Tc-99m throughout the pharmacy.
The incident resulted in the contamination of pharmacy staff, as well as packages prepared for shipment to pharmacy customers. Some of the packages were shipped to customers with levels of removable contamination that exceeded regulatory limits. The incident did not result in significant external or internal dose to pharmacy staff.
Discussion:
Volatilization of radiochemicals or radiopharmaceuticals can create an airborne hazard, potentially resulting in internal doses to workers and spread of contamination to unrestricted areas through ventilation systems.
Licensees should review their procedures for handling radioactive materials to identify processes that could cause volatilization. In addition to direct heating, as in the example above, exothermic chemical reactions and changes in pH can result in volatilization of some materials. Some organic compounds may volatilize at room temperature if stored in open containers.
Licensees should ensure that their emergency procedures adequately address this scenario if they handle or process radioactive materials in a manner that could cause volatilization. The emergency procedures should include, at a minimum, instructions to immediately evacuate and secure the affected areas. Licensees should also assess their ventilation systems and determine whether emergency procedures are appropriate to prevent circulation of radioactive contamination to other areas within the facility. Licensees may also consider performing heating procedures or potential volatile processes in a laboratory hood with independent ventilation.
This information notice requires no specific action nor written response. If you have any questions about the information in this notice, please contact the technical contact listed below or the appropriate regional office.