Home > NRC Library > Document Collections > General Communications > Information Notices > 1988 > IN 88-52
UNITED STATES
NUCLEAR REGULATORY COMMISSION
OFFICE OF NUCLEAR MATERIAL SAFETY AND SAFEGUARDS
WASHINGTON, D.C. 20555
July 27, 1988
Information Notice No. 88-52: FAILURE OF INTRAUTERINE TANDEM OF FLETCHER
SUIT APPLICATOR BRACHYTHERAPY DEVICES
DURING PATIENT TREATMENT
Addressees:
Medical licensees.
Purpose:
This information notice is intended to provide information on the inspection
of intrauterine tandem of Fletcher Suit Applicator brachytherapy devices for
applicator weld corrosion and the proper sterilization procedures. It is ex-
pected that licensees will review this information for applicability to their
brachytherapy activities and distribute this notice to responsible medical
staff. Suggestions contained in this information notice do not constitute new
NRC requirements, and no written response is required.
Description of Circumstances:
The U.S. Food and Drug Administration (FDA) Center for Devices and
Radiological Health received eight reports since 1983 of incidents with
Fletcher Suit Applicator intrauterine tandems. In each incident, the
radioactive capsules were removed, but the closed end of the tandems separated
from the applicator and remained in the patient during applicator removal.
Additional radiographic and operating procedures were required to locate and
remove the broken pieces. The failure is caused by weld joint corrosion. The
corrosion may be due to improper cleaning before sterilization.
Discussion:
The Fletcher Suit Applicators can be used with either byproduct material or
radium brachytherapy sources. The FDA believes that all Fletcher Suit Ap-
plicators should be inspected before each use, to ensure that the closed ends
of the intrauterine tandem are secure. If they are not secure, the device
should not be used. Cleaning the applicators with hydrogen peroxide or
glutaraldehyde (Cydex) before autoclaving or cold sterilization may corrode
the applicators' welds. The FDA recommends that the applicators be cleaned
using only mild detergent and water, and then thoroughly rinsed with water
before sterilization. The FDA also recommends that all applicators of similar
design be checked, and the same cleaning instructions followed.
8807210218
. IN 88-52
July 27, 1988
Page 2 of 2
No specific action or written response is required by this information notice.
If you have any questions about this matter, please contact the person listed
below or the appropriate NRC regional office.
Richard E. Cunningham, Director
Division of Industrial and
Medical Nuclear Safety
Office of Nuclear Material Safety
and Safeguards
Technical Contact: Donna-Beth Howe, NMSS
(301) 492-0636
Attachment: List of Recently Issued NRC Information Notices
. Attachment
IN 88-52
July 27, 1988
Page 1 of 1
LIST OF RECENTLY ISSUED
NRC INFORMATION NOTICES
_____________________________________________________________________________
Information Date of
Notice No._____Subject_______________________Issuance_______Issued to________
88-46, Licensee Report of 7/26/88 All holders of OLs
Supplement 1 Defective Refurbished or CPs for nuclear
Circuit Breakers power reactors.
88-51 Failures of Main Steam 7/21/88 All holders of OLs
Isolation Valves or CPs for nuclear
power reactors.
88-50 Effect of Circuit 7/18/88 All holders of OLs
Breaker Capacitance or CPs for nuclear
on Availability of power reactors.
Emergency Power
88-49 Marking, Handling, 7/18/88 All holders of OLs
Control, Storage and or CPs for nuclear
Destruction of Safe- power reactors and
guards Information all other licensed
activities
involving a
formula quantity
of special nuclear
material.
88-48 Licensee Report of 7/12/88 All holders of OLs
Defective Refurbished or CPs for nuclear
Valves power reactors.
88-47 Slower-Than-Expected 7/14/88 All holders of OLs
Rod-Drop Times or CPs for PWRs.
88-46 Licensee Report of 7/8/88 All holders of OLs
Defective Refurbished or CPs for nuclear
Circuit Breakers power reactors.
88-45 Problems In Protective 7/7/88 All holders of OLs
Relay and Circuit or CPs for nuclear
Breaker Coordination power reactors.
88-44 Mechanical Binding of 6/24/88 All holders of OLs
Spring Release Device or CPs for nuclear
in Westinghouse Type power reactors.
DS-416 Circuit Breakers
_____________________________________________________________________________
OL = Operating License
CP = Construction Permit
Page Last Reviewed/Updated Thursday, March 29, 2012