U.S. Nuclear Regulatory Commission
Technical Assistance Request, Clarification of 10 CFR 35.50 (b) (1)
HPPOS-280 PDR-9306150132
Title: Technical Assistance Request, Clarification of 10
CFR 35.50 (b) (1)
See the memorandum from J.E. Glenn to W. E. Cline dated
November 12, 1991. This memo responds to a Region II
request for clarification of the performance of dose
calibrator consistency checks as described in 10 CFR 35.50
(b) (1). Specifically, "is it appropriate for a licensee to
preform this test on the cobalt-57 (Co-57) setting although
technetium-99m (Tc-99m) is the most frequently used
isotope?"
Medical licensees are required to perform a dose calibrator
daily constancy check. 10 CFR 35.50 (b) (1) requires, in
part, that a licensee check each dose calibrator for
constancy with a dedicated check source at the beginning of
each day of use, and that the test be done on a frequently
used setting. Based on numerous nuclear medicine
inspections in Region II, the most frequently used setting
is Tc-99m, and based on the requirements of 10 CFR 35.51
(b) (1), the licensees who use Tc-99m more frequently
should perform this test on the Tc-99m setting with a
dedicated sealed source (which is usually Co-57). However,
some licensees perform this test on the Co-57 setting
although Tc-99m is the most frequently used setting.
While this issue is not addressed in the Statements of
Consideration for either the proposed or final rule on 10
CFR Part 35, effective April 1, 1987, it is believed that
the rule is intended to assure that the licensee determines
the consistency of the dose calibrator, on each day of use,
under the actual conditions of use. Since most medical
licensees use Tc-99m for patient dosage administrations
more frequently than any other isotope, such licensees must
check the Tc-99m setting, on each day of use, with a
dedicated check source. If the licensee frequently uses
the Tc-99m setting to measure patient dosages but only does
a constancy check on the Co-57 setting, it appears
appropriate to cite against 10 CFR 35.50 (b) (1) unless the
licensee can show that the Co-57 setting is frequently used
to measure patient dosages.
It is recommended that Co-57 be used as a standard to
measure the constancy of the Tc-99m setting because of the
close proximity of its energies. Cobalt-57 has principal
energies of 122 and 136 keV and Tc-99m has a principal
energy of 140 keV. It is also recommended that dose
calibrators having pre-calibrated settings or
potentiometers be tested on both the Co-57 and Tc-99m
settings because discrepancies or fluctuations have been
observed between the two settings when tested for constancy
with the same check source. If such discrepancies are
observed, it could indicate that there is a problem with
one or both of the settings. Inspectors should encourage
licensees to do a daily constancy check of all commonly
used isotope settings, not only Tc-99m to ensure the
accuracy of all administered patient dosages.
Regulatory references: 10 CFR 35.50
Subject codes: 6.4, 6.6
Applicability: Byproduct Materials

