U.S. Nuclear Regulatory Commission
Updated Guidance on Fit Testing of Biopak 60-P Respirator Users
HPPOS-146 PDR-9111210387
Title: Updated Guidance on Fit Testing of Biopak 60-P
Respirator Users
See the memorandum from L. J. Cunningham to J. N. Grace
(and others) dated August 19, 1984. It is acceptable to
check the fit of the Biopak 60-P while the user is wearing
just the facepiece equipped with a high efficiency filter
supplied by the manufacturer of the device. A fit factor
of 1000 is reasonable in the negative pressure
air-purifying mode. This memo supersedes an earlier one
from L. J. Cunningham to L. R. Greger dated August 8, 1983.
Licensees and inspectors had inquired as to what
constitutes an acceptable method for performing
quantitative fitting of the wearers of this apparatus as
required in footnote 1, to Appendix A of Part 20.
Specifically, was it acceptable to check the fit of the
device (the face to facepiece sealing capability) by
testing the user while the user was wearing just the
facepiece equipped with a high efficiency filter supplied
by the manufacturer of the device?
Previous guidance stated that the wearer must don the
entire unit for fit testing since it was felt that fitting
the facepiece with a high efficiency filter that is capable
of allowing no more than 0.03% leakage would preclude
measurement of the required 0.02% leakage or less through
the face to facepiece sealing area. However, the 0.03%
leakage allowed for high efficiency filters is determined
with a more penetrating aerosol (monodispersed) than used
in fit testing. Therefore, it is possible to measure the
0.02% leakage accurately with the facepiece equipped with a
high efficiency filter (0.02% leakage corresponds to a fit
factor of 5000).
Requiring a fit factor of 5000 in the negative pressure
air-purifying mode is too restrictive. This approach to
fit testing allows no credit for protection provided by the
positive pressure inside the facepiece generated by the
device in its normal mode of operation. Positive pressure
inside the facepiece can compensate for inward leakage of
contaminants to some extent by ensuring air circulating
through the device is leaked outward instead of leaking
contaminants into the worker's breathing zone. However,
with this device, protection is obtained at a large cost if
the fit is poor and outward leakage is substantial because
reduced service life results as outward leakage of air is
made up from the small volume of oxygen carried by the
user. The volume carried is sufficient to exchange the
volume of carbon dioxide released in respiration with
compressed oxygen. Carbon dioxide is removed from the
circulating air by the sorbent scrubber.
A hard and fast number that delineates good from poorly
fitting respirators is not available. In the opinion of
many experts in the field of respiratory protection, a fit
factor of 1000 seems reasonable for distinguishing between
good and poorly fitting respirators. It is recommended
that licensees use this number as a guide for determining
if an acceptable fit has been achieved with this device.
For those persons who are unable to attain a fit factor of
1000 with just the facepiece in negative pressure mode,
participation in emergency, potentially IDLH situations
should be restricted. This person may experience
drastically reduced service time which reduces emergency
response capability as well as hindering escape from a
potentially life threatening situation.
The intent of the previous guidance was not to verify
proper functioning of the entire unit. The operability of
the assembled unit is checked after maintenance and before
each use. In addition, fit testing of workers wearing the
assembled unit in the case of this apparatus was presenting
other problems due to the low makeup volume and leakage
detection interference from background water vapor droplets
and particulates from the carbon dioxide scrubber system.
Based on the interference problem that has been reported
and reevaluation of the previous guidance, it is
recommended that fit testing of wearers of the BioPak 60-P
be performed with just the facepiece equipped with a high
efficiency filter and that a factor of 1000 be considered
an acceptable fit. A recommendation will be made to RES to
update Appendix A to include the intent of this
interpretation in the next rule change.
Regulatory references: 10 CFR 20.103, 10 CFR 20.1703,
Regulatory Guide 8.15
Subject codes: 7.2, 8.10
Applicability: All

